| Literature DB >> 15328405 |
Hongkai Wu1, Aaron Wheeler, Richard N Zare.
Abstract
An integrated microfluidic device has been fabricated for analyzing the chemical contents of a single cell (chemical cytometry). The device is designed to accomplish four different functions: (i) cell handling, (ii) metering and delivering of chemical reagents, (iii) cell lysis and chemical derivatization, and (iv) separating derivatized compounds and detecting them by laser-induced fluorescence. These functions are accomplished with only two valves, formed by multilayer soft lithography. A new kind of three-state valve and a picopipette are described; these elements are crucial for minimizing the reaction volume and ensuring optimal shape of the channel for electrophoresis injection. By using these valves, a reaction volume of approximately 70 pl is achieved for the lysis and derivitization of the contents of a single Jurkat T cell (approximately 10 microm diameter). As a demonstration of the use of this integrated microfluidic device, electropherograms of amino acids from individual Jurkat T cells are recorded and compared with those collected from a multiple-cell homogenate. Copyright 2004 The National Academy of Sciencs of the USAMesh:
Year: 2004 PMID: 15328405 PMCID: PMC516477 DOI: 10.1073/pnas.0405299101
Source DB: PubMed Journal: Proc Natl Acad Sci U S A ISSN: 0027-8424 Impact factor: 11.205