[Transpalpebral tonometry with a digital tonometer in healthy eyes and after penetrating keratoplasty].

M Amm1, J Hedderich.   

Abstract

PURPOSE: The Goldmann applanation technique established itself as gold standard for clinical intraocular pressure (IOP) measurement. But the device measures by corneal touch and presupposes corneal conditions within a normal range. So precise determination of IOP is crucial in corneal disorders and after corneal surgery. This pilot study presents tonometry data of normal and postkeratoplasty eyes measured transpalpebrally with a digital tonometer in comparison to Goldmann tonometry.
METHOD: The TGDc-01 tonometer is based on a ballistic principle of measurement. IOP was measured in 50 healthy eyes (group 1) and 40 eyes after penetrating keratoplasty (group 2) first by the TGDc-01 and after 3 minutes by Goldmann tonometry (Haag-Streit). Subjective refraction, keratometry and ultrasonic pachymetry were performed in all eyes. The IOP measurements in the postkeratoplasty eyes were done when corneal epithelium was fully closed and only a medium suture edema could be observed.
RESULTS: The Pearson's product-moment correlation was r=0,46 for group 1 and 0,65 for group 2. Using the concordance correlation coefficient the value showed a reproducibility index of r(c)=0,44 in group 1 and r(c)=0,60 in group 2. Following the Bland-Altman-analysis 58% of the cases in group 1 had a difference between the two instruments of maximum 3 mmHg. In group 2, 53,5% of the eyes showed an absolute difference between Goldmann and TDGc-01 of 3 mmHg or less.
CONCLUSION: A transpalpebral measurement technique-avoiding corneal contact-seems to be an alternative choice in patients with corneal pathologies and after corneal surgery. The TGDc-01 tonometer provided mostly similar results compared to the Goldmann device, but with fluctuations. Keeping the deviations between the two methods in mind we recommend, however, the use of the transpalpebral approach only in combination with other techniques (palpation, indentation) to obtain enough reliability of IOP information in problem eyes.

Entities:  

Mesh:

Year:  2005        PMID: 15322801     DOI: 10.1007/s00347-004-1082-5

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  10 in total

1.  IOP measurement and central corneal thickness.

Authors:  Richard Stodtmeister; Martina Kron; Wilhelm Gaus
Journal:  Br J Ophthalmol       Date:  2002-01       Impact factor: 4.638

2.  [Impression and applanation tonometry in irregular corneas. Comparison with intraocular needle tonometry].

Authors:  F Madjlessi; W Marx; T Reinhard; C Althaus; R Sundmacher
Journal:  Ophthalmologe       Date:  2000-07       Impact factor: 1.059

3.  Accurate intraocular pressure measurement-the myth of modern ophthalmology?

Authors:  S Shah
Journal:  Ophthalmology       Date:  2000-10       Impact factor: 12.079

4.  Applanation tonometry and central corneal thickness.

Authors:  N Ehlers; T Bramsen; S Sperling
Journal:  Acta Ophthalmol (Copenh)       Date:  1975-03

5.  Central corneal thickness on GAT (Goldman applanation tonometry accuracy).

Authors:  Richard Stodtmeister
Journal:  J Glaucoma       Date:  2002-12       Impact factor: 2.503

Review 6.  Influence of corneal variables on accuracy of intraocular pressure measurement.

Authors:  Karim F Damji; Rajeev H Muni; Rejean M Munger
Journal:  J Glaucoma       Date:  2003-02       Impact factor: 2.503

7.  [A new applanation tonometer].

Authors:  H GOLDMANN
Journal:  Bull Mem Soc Fr Ophtalmol       Date:  1954

8.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

9.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 10.  The non-contact tonometer. Its value and limitations.

Authors:  M B Shields
Journal:  Surv Ophthalmol       Date:  1980 Jan-Feb       Impact factor: 6.048

  10 in total
  9 in total

1.  Deviations between transpalpebral tonometry using TGDc-01 and Goldmann applanation tonometry depending on the IOP level.

Authors:  Annette Troost; Kathrin Specht; Frank Krummenauer; Sung Hyun Yun; Oliver Schwenn
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-15       Impact factor: 3.117

2.  Effects of corneal thickness on intraocular pressure measured with three different tonometers.

Authors:  Hiroki Murase; Akira Sawada; Kiyofumi Mochizuki; Tetsuya Yamamoto
Journal:  Jpn J Ophthalmol       Date:  2009-01-30       Impact factor: 2.447

Review 3.  [Measuring intraocular pressure by different methods].

Authors:  J Lamparter; E M Hoffmann
Journal:  Ophthalmologe       Date:  2009-08       Impact factor: 1.059

4.  Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers.

Authors:  A Troost; S H Yun; K Specht; F Krummenauer; O Schwenn
Journal:  Br J Ophthalmol       Date:  2005-03       Impact factor: 4.638

5.  [Glaucoma attack following Hughes tarsoconjunctival flap: what should be done?].

Authors:  E Bertelmann; N Minko; N Torun
Journal:  Ophthalmologe       Date:  2012-09       Impact factor: 1.059

6.  Central corneal thickness and Diaton transpalpebral tonometry.

Authors:  Mustafa Ilker Toker; Ayse Vural; Haydar Erdogan; Aysen Topalkara; Mustafa Kemal Arici
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-02-20       Impact factor: 3.117

Review 7.  Graft failure: III. Glaucoma escalation after penetrating keratoplasty.

Authors:  Emily C Greenlee; Young H Kwon
Journal:  Int Ophthalmol       Date:  2008-06       Impact factor: 2.031

8.  Dual Optical Signal-based Intraocular Pressure-sensing Principle Using Pressure-sensitive Mechanoluminescent ZnS:Cu/PDMS Soft Composite.

Authors:  Yooil Kim; Sunanda Roy; Gwang-Yong Jung; Jung-Sik Oh; Gi-Woo Kim
Journal:  Sci Rep       Date:  2019-10-23       Impact factor: 4.379

9.  Utilizing the principles of Gagne's nine events of instruction in the teaching of Goldmann Applanation Tonometry.

Authors:  Yee Ling Wong
Journal:  Adv Med Educ Pract       Date:  2018-01-11
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.