Literature DB >> 15321934

Spectral entropy measurement of patient responsiveness during propofol and remifentanil. A comparison with the bispectral index.

A L G Vanluchene1, M M R F Struys, B E K Heyse, E P Mortier.   

Abstract

BACKGROUND: We compared two spectral entropies, state entropy (SE) and response entropy (RE), based on the irregularity of the EEG, to measure loss of response to verbal command (LOR(verbal)) and noxious stimulus (LOR(noxious)) with the bispectral index (BIS) during propofol infusion with and without remifentanil.
METHODS: Three groups of 20 patients received an effect-site controlled propofol infusion (Ce(PROP)) starting at 1 microg ml(-1) and increased in steps of 0.5 microg ml(-1) at 4 min intervals. In addition, a remifentanil infusion was maintained at a group-dependent, fixed effect-site target concentration (Ce(REMI)) (0, 2 or 4 ng ml(-1)). The ability of BIS, SE or RE to predict LOR(verbal) and LOR(noxious) were compared with the changes in BIS, SE and RE using logistic regression, prediction probability (P(K)), and sensitivity/specificity.
RESULTS: In all groups, BIS, SE and RE decreased with increasing Ce(PROP). However, BIS decreased more smoothly than SE and RE at deeper levels of sedation. At LOR(verbal), BIS(50), SE(50) and RE(50) increased with increasing Ce(REMI). BIS, SE and RE all detected LOR(verbal) accurately but BIS performed better at 100% sensitivity. Sensitivity/specificity for detection of LOR(verbal) decreased for all methods with increasing Ce(REMI). LOR(noxious) was poorly described by all measures.
CONCLUSION: LOR(verbal) was detected accurately by BIS, SE and RE except for 100% sensitivity, where BIS performed better. Though BIS, SE and RE were influenced by remifentanil during propofol administration, their ability to detect LOR(verbal) remained accurate. None of the measures predicted LOR(noxious).

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15321934     DOI: 10.1093/bja/aeh251

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  19 in total

1.  Response entropy changes after noxius stimulus.

Authors:  Jose L Guerrero; E Matute; E Alsina; B Del Blanco; F Gilsanz
Journal:  J Clin Monit Comput       Date:  2012-03-20       Impact factor: 2.502

Review 2.  Using EEG to monitor anesthesia drug effects during surgery.

Authors:  Leslie C Jameson; Tod B Sloan
Journal:  J Clin Monit Comput       Date:  2006-12       Impact factor: 2.502

Review 3.  Mechanisms of anesthetic actions and the brain.

Authors:  Yumiko Ishizawa
Journal:  J Anesth       Date:  2007-05-30       Impact factor: 2.078

Review 4.  [Measurement of the depth of anaesthesia].

Authors:  G N Schmidt; J Müller; P Bischoff
Journal:  Anaesthesist       Date:  2008-01       Impact factor: 1.041

5.  Response entropy is more reactive than bispectral index during laparoscopic gastric banding.

Authors:  James Feld; William E Hoffman
Journal:  J Clin Monit Comput       Date:  2006-07-04       Impact factor: 2.502

6.  Comparison of spectral entropy and BIS VISTA™ monitor during general anesthesia for cardiac surgery.

Authors:  Tadeusz Musialowicz; Pasi Lahtinen; Otto Pitkänen; Jouni Kurola; Ilkka Parviainen
Journal:  J Clin Monit Comput       Date:  2011-04-22       Impact factor: 2.502

7.  Auditory Predictive Coding across Awareness States under Anesthesia: An Intracranial Electrophysiology Study.

Authors:  Kirill V Nourski; Mitchell Steinschneider; Ariane E Rhone; Hiroto Kawasaki; Matthew A Howard; Matthew I Banks
Journal:  J Neurosci       Date:  2018-08-20       Impact factor: 6.167

8.  Cortical functional connectivity indexes arousal state during sleep and anesthesia.

Authors:  Matthew I Banks; Bryan M Krause; Christopher M Endemann; Declan I Campbell; Christopher K Kovach; Mark Eric Dyken; Hiroto Kawasaki; Kirill V Nourski
Journal:  Neuroimage       Date:  2020-02-08       Impact factor: 6.556

9.  Spectral entropy for assessing the depth of propofol sedation.

Authors:  Mi-Young Kwon; Seung-Yun Lee; Tae-Yop Kim; Duk Kyung Kim; Kyoung-Min Lee; Nam-Sik Woo; Young-Jae Chang; Myung Ae Lee
Journal:  Korean J Anesthesiol       Date:  2012-03-21

Review 10.  Spectral entropy monitoring for adults and children undergoing general anaesthesia.

Authors:  Anjolie Chhabra; Rajeshwari Subramaniam; Anurag Srivastava; Hemanshu Prabhakar; Mani Kalaivani; Saloni Paranjape
Journal:  Cochrane Database Syst Rev       Date:  2016-03-14