Effect of prosthetic aortic valve design on the Doppler-catheter gradient correlation: an in vitro study of normal St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves.

H Baumgartner1, S Khan, M DeRobertis, L Czer, G Maurer.   

Abstract

To evaluate the normal range of Doppler-derived velocities and gradients, their relation to direct flow measurements and the importance of prosthetic valve design on the relation between Doppler and catheter-derived gradients, five sizes of normal St. Jude bileaflet, Medtronic-Hall tilting disc, Starr-Edwards caged ball and Hancock bioprosthetic aortic valves were studied with use of a pulsatile flow model. A strong linear correlation between peak velocity and peak flow, and mean velocity and mean flow, was found in all four valve types (r = 0.96 to 0.99). In small St. Jude and Hancock valves, Doppler velocities and corresponding gradients increased dramatically with increasing flow, resulting in velocities and gradients as high as 4.7 m/s and 89 mm Hg, respectively. The ratio of velocity across the valve to velocity in front of the valve (velocity ratio) was independent of flow in all St. Jude, Medtronic-Hall, Starr-Edwards and Hancock valves when the two lowest flow rates were excluded for Hancock valves. Although Doppler peak and mean gradients correlated well with catheter peak and mean gradients in all four valve types, the actual agreement between the two techniques was acceptable only in Hancock and Medtronic-Hall valves. For St. Jude and Starr-Edwards valves, Doppler gradients significantly and consistently exceeded catheter gradients with differences as great as 44 mm Hg. Thus, Doppler velocities and gradients across normal prosthetic heart valves are highly flow dependent. However, the velocity ratio is independent of flow.(ABSTRACT TRUNCATED AT 250 WORDS)

Mesh:

Year:  1992        PMID: 1531058     DOI: 10.1016/0735-1097(92)90486-7

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  7 in total

1.  A novel method of assessing Starr-Edwards aortic valve stenosis.

Authors:  R J Edwards; M Al-Bustami; C D J Ilsley
Journal:  Heart       Date:  2004-12       Impact factor: 5.994

2.  Discrepancy between Doppler and catheter measurements of pressure gradients across small-size prosthetic valve.

Authors:  Takuya Yamashita; Yukinori Moriyama; Naoyuki Sata; Naokazu Hamada; Takashi Horinouchi; Shigeru Amitani; Kenkichi Miyahara; Kentaro Setoyama; Kazuhiro Misumi; Hiroshi Sakamoto
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2005-02

Review 3.  Stenotic lesions.

Authors:  B Wranne; H Baumgartner; F Flachskampf; M Hasenkam; F Pinto
Journal:  Heart       Date:  1996-06       Impact factor: 5.994

4.  Detection of "pathologic" prosthetic valve stenosis via exercise Doppler echocardiography.

Authors:  M Shigenobu; S Sano
Journal:  Heart Vessels       Date:  1995       Impact factor: 2.037

5.  Discrepancies between Doppler and catheter gradients in ventricular septal defect: a correction of localized gradients from pressure recovery phenomenon.

Authors:  Taylan Akgun; Can Yücel Karabay; Gonenc Kocabay; Vecih Oduncu; Arzu Kalayci; Ahmet Guler; Olcay Ozveren; Fatih Yilmaz; Mustafa Akcakoyun; Cevat Kirma
Journal:  Int J Cardiovasc Imaging       Date:  2013-09-14       Impact factor: 2.357

6.  Invasive versus echocardiographic gradients in degenerated surgical aortic valve prostheses: A multicenter study.

Authors:  Amer N Kadri; George Hanzel; Sammy Elmariah; Francis Shannon; Karim Al-Azizi; Judith Boura; Michael Mack; Amr E Abbas
Journal:  JTCVS Open       Date:  2021-07-17

7.  Comparison of Transvalvular Aortic Mean Gradients Obtained by Intraprocedural Echocardiography and Invasive Measurement in Balloon and Self-Expanding Transcatheter Valves.

Authors:  Amr E Abbas; Ramy Mando; Amer Kadri; Houman Khalili; George Hanzel; Francis Shannon; Karim Al-Azizi; Thomas Waggoner; Safwan Kassas; Thomas Pilgrim; Taishi Okuno; Alexander Camacho; Alexandra Selberg; Sammy Elmariah; Anthony Bavry; Julien Ternacle; Jared Christensen; Neil Gheewala; Philippe Pibarot; Michael Mack
Journal:  J Am Heart Assoc       Date:  2021-09-29       Impact factor: 5.501