Dose verification in clinical IMRT prostate incidents.

Panagiotis Sandilos1, Angelos Angelopoulos, Panagiotis Baras, Konstantinos Dardoufas, Pantelis Karaiskos, Panagiotis Kipouros, Marek Kozicki, Janusz M Rosiak, Loukas Sakelliou, Ioannis Seimenis, Lampros Vlahos.   

Abstract

PURPOSE: In view of the need for dose-validation procedures on each individual intensity-modulated radiation therapy (IMRT) plan, dose-verification measurements by film, by ionization chamber, and by polymer gel-MRI dosimetry were performed for a prostate-treatment plan configuration. Treatment planning system (TPS) calculations were evaluated against dose measurements. METHODS AND MATERIALS: Intensity-modulated radiation therapy (IMRT) treatments were planned on a commercial TPS. Kodak EDR-2 films were used for the verification of two-dimensional (2D) dose distributions at 1 coronal and 5 axial planes in a water-equivalent phantom. Full three-dimensional (3D) dose distributions were measured by use of a novel polymer gel formulation and a 3D magnetic resonance imaging (MRI) readout technique. Calculations were compared against measurements by means of isocontour maps, gamma-index maps (3% dose difference, 3-mm distance to agreement) and dose-volume histograms.
RESULTS: A good agreement was found between film measurements and TPS predictions for points within the 60% isocontour, for all the examined plans (gamma-index <1 for 96% of pixels). Three-dimensional dose distributions obtained with the polymer gel-MRI method were adequately matched with corresponding TPS calculations, for measurements in a gel phantom covering the planning-target volume (PTV).
CONCLUSIONS: Measured 2D and 3D dose distributions suggest that, for the investigated prostate IMRT plan configuration, TPS calculations provide clinically acceptable accuracy.

Entities:  

Mesh:

Year:  2004        PMID: 15275742     DOI: 10.1016/j.ijrobp.2004.04.029

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  8 in total

1.  A novel technique to enable experimental validation of deformable dose accumulation.

Authors:  Carolyn J Niu; Warren D Foltz; Michael Velec; Joanne L Moseley; Adil Al-Mayah; Kristy K Brock
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

2.  Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using presage/optical-CT.

Authors:  Harshad Sakhalkar; David Sterling; John Adamovics; Geoffrey Ibbott; Mark Oldham
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

3.  Evaluating deviations in prostatectomy patients treated with IMRT.

Authors:  Ana Cravo Sá; Ana Peres; Mónica Pereira; Carina Marques Coelho; Fátima Monsanto; Ana Macedo; Adrian Lamas
Journal:  Rep Pract Oncol Radiother       Date:  2015-12-29

Review 4.  Polymer gel dosimetry.

Authors:  C Baldock; Y De Deene; S Doran; G Ibbott; A Jirasek; M Lepage; K B McAuley; M Oldham; L J Schreiner
Journal:  Phys Med Biol       Date:  2010-02-11       Impact factor: 3.609

5.  Acrylonitrile Butadiene Styrene (ABS) plastic based low cost tissue equivalent phantom for verification dosimetry in IMRT.

Authors:  Rajesh Kumar; S D Sharma; Sudesh Deshpande; Yogesh Ghadi; V S Shaiju; H I Amols; Y S Mayya
Journal:  J Appl Clin Med Phys       Date:  2009-12-17       Impact factor: 2.102

Review 6.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19

7.  Mechanical and dose delivery accuracy evaluation in radiosurgery using polymer gels.

Authors:  Panagiotis Sandilos; Elias Tatsis; Lampros Vlachos; Pantelis Karaiskos; Evangelos Georgiou; Panagiotis Kipouros; Michael Torrens; Angelos Angelopoulos
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

8.  Image guided IMRT dosimetry using anatomy specific MOSFET configurations.

Authors:  Md Nurul Amin; Bern Norrlinger; Robert Heaton; Mohammad Islam
Journal:  J Appl Clin Med Phys       Date:  2008-06-23       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.