Literature DB >> 15220498

Patient radiation dose at CT urography and conventional urography.

Richard D Nawfel1, Philip F Judy, A Robert Schleipman, Stuart G Silverman.   

Abstract

PURPOSE: To measure and compare patient radiation dose from computed tomographic (CT) urography and conventional urography and to compare these doses with dose estimates determined from phantom measurements.
MATERIALS AND METHODS: Patient skin doses were determined by placing a thermoluminescent dosimeter (TLD) strip (six TLD chips) on the abdomen of eight patients examined with CT urography and 11 patients examined with conventional urography. CT urography group consisted of two women and six men (mean age, 55.5 years), and conventional urography group consisted of six women and five men (mean age, 58.9 years). CT urography protocol included three volumetric acquisitions of the abdomen and pelvis. Conventional urography protocol consisted of acquisition of several images involving full nephrotomography and oblique projections. Mean and SD of measured patient doses were compared with corresponding calculated doses and with dose measured on a Lucite pelvic-torso phantom. Correlation coefficient (R(2)) was calculated to compare measured and calculated skin doses for conventional urography examination, and two-tailed P value significance test was used to evaluate variation in effective dose with patient size. Radiation risk was calculated from effective dose estimates.
RESULTS: Mean patient skin doses for CT urography measured with TLD strips and calculated from phantom data (CT dose index) were 56.3 mGy +/- 11.5 and 54.6 mGy +/- 4.1, respectively. Mean patient skin doses for conventional urography measured with TLD strips and calculated as entrance skin dose were 151 mGy +/- 90 and 145 mGy +/- 76, respectively. Correlation coefficient between measured and calculated skin doses for conventional urography examinations was 0.95. Mean effective dose estimates for CT urography and conventional urography were 14.8 mSv +/- 90.0 and 9.7 mSv +/- 3.0, respectively. Mean effective doses estimated for the pelvic-torso phantom were 15.9 mSv (CT urography) and 7.8 mSv (conventional urography).
CONCLUSION: Standard protocol for CT urography led to higher mean effective dose, approximately 1.5 times the radiation risk for conventional urography. Patient dose estimates should be taken into consideration when imaging protocols are established for CT urography. Copyright RSNA, 2004

Entities:  

Mesh:

Year:  2004        PMID: 15220498     DOI: 10.1148/radiol.2321030222

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  32 in total

1.  Current status of low dose multi-detector CT in the urinary tract.

Authors:  Mi Kim Sung; Sarabjeet Singh; Mannudeep K Kalra
Journal:  World J Radiol       Date:  2011-11-28

Review 2.  Obstructive diseases of the urinary tract in children: lessons from the last 15 years.

Authors:  Michael Riccabona
Journal:  Pediatr Radiol       Date:  2010-04-30

3.  Multi-detector CT urography: effect of oral hydration and contrast medium volume on renal parenchymal enhancement and urinary tract opacification--a quantitative and qualitative analysis.

Authors:  Dieter H Szolar; Manfred Tillich; Klaus W Preidler
Journal:  Eur Radiol       Date:  2010-04-10       Impact factor: 5.315

Review 4.  [Multislice CT urography Aspects for technical management and clinical application].

Authors:  J Kemper; G Adam; C Nolte-Ernsting
Journal:  Radiologe       Date:  2005-10       Impact factor: 0.635

5.  Dose reduction in multidetector CT of the urinary tract. Studies in a phantom model.

Authors:  E Coppenrath; T Meindl; P Herzog; R Khalil; U Mueller-Lisse; L Krenn; M Reiser; U G Mueller-Lisse
Journal:  Eur Radiol       Date:  2006-03-28       Impact factor: 5.315

6.  MDCT urography: retrospective determination of optimal delay time after intravenous contrast administration.

Authors:  Thomas Meindl; Eva Coppenrath; Rami Kahlil; Ulrike L Müller-Lisse; Maximilian F Reiser; Ullrich G Müller-Lisse
Journal:  Eur Radiol       Date:  2006-04-01       Impact factor: 5.315

Review 7.  Understanding multislice CT urography techniques: Many roads lead to Rome.

Authors:  Claus Nolte-Ernsting; Nigel Cowan
Journal:  Eur Radiol       Date:  2006-09-05       Impact factor: 5.315

Review 8.  CT urography: the end of IVU?

Authors:  F Stacul; A Rossi; M A Cova
Journal:  Radiol Med       Date:  2008-06-06       Impact factor: 3.469

9.  Unenhanced helical CT following excretory urography in the diagnosis of upper urinary tract disease: a little more cost, a lot more value.

Authors:  Hong Hu; Xiao-Yun Hu; Xiang-Ming Fang; Hong-Wei Chen; Xia-Juan Yao
Journal:  Urol Res       Date:  2009-11-27

10.  Analysis of 777 cases with obstruction of the ureter or extrahepatic bile duct by ultrasonography after normal saline retention enema.

Authors:  Chong Tang; Xuegang Wu; Qiuhong Fan; Zhensheng Deng
Journal:  Crit Ultrasound J       Date:  2012-04-17
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.