Literature DB >> 15216292

ECHO-PHONOCARDIOGRAPHIC CHARACTERISTICS OF THE NORMALLY FUNCTIONING COOLEY-CUTTER MITRAL VALVE PROSTHESIS.

Leonard W. Pechacek1, Ali Zarrabi, Ali Massumkhani, Efrain Garcia, Carlos M. De Castro, Robert J. Hall.   

Abstract

Echocardiographic and phonocardiographic records of 19 patients with a normally functioning Cooley-Cutter mitral valve were analyzed in order to provide quantitative baseline values for this prosthesis. The average duration between the second heart sound and peak opening of the valve (A(2)-OC interval) was 83 +/- 4 (standard error of the mean) msec. The Q-CC interval (from the electrocardiographic Q wave to closure of the valve) was 71 +/- 2 msec. Mean opening and closing velocities of the disc were similar (396 +/- 11 mm/sec and 393 +/- 12 mm/sec, respectively). Amplitude of disc excursion ranged from 6 to 9 mm, but this measurement was not possible in all patients due to the presence of spurious echoes. Early onset of prosthetic valve closure was a relatively common finding in patients with atrial fibrillation or with various forms of atrioventricular block, and frequently resulted in a variety of phonocardiographic alterations. Except for the A(2)-OC interval and amplitude of disc excursion, there was no significant correlation between valve size and echo-phonocardiographic measurements. Results of this study are compared with values previously reported for other types of caged disc valves, and the usefulness and limitations of echo-phonocardiographic assessment of prosthetic valve function are briefly discussed.

Entities:  

Year:  1979        PMID: 15216292      PMCID: PMC287814     

Source DB:  PubMed          Journal:  Cardiovasc Dis        ISSN: 0093-3546


  14 in total

Review 1.  Resolution problems in echocardiology: a source of interpretation errors.

Authors:  J Roelandt; W G van Dorp; N Bom; J D Laird; P G Hugenholtz
Journal:  Am J Cardiol       Date:  1976-02       Impact factor: 2.778

2.  Ultrasonic measurement of prosthetic heart valve action.

Authors:  G McIlmoyle; D A Robertson; W N McDicken; D H Evans
Journal:  Cardiovasc Res       Date:  1975-07       Impact factor: 10.787

3.  Delayed opening of Beall mitral prosthetic valve detected by echocardiography.

Authors:  N Kawai; B L Segal; J W Linhart
Journal:  Chest       Date:  1975-02       Impact factor: 9.410

4.  Echocardiographic and phonocardiographic confirmation of suspected caged mitral valve malfunction.

Authors:  T B Berndt; D J Goodman; R L Popp
Journal:  Chest       Date:  1976-08       Impact factor: 9.410

5.  Malfunction of a Cutter-Smeloff mitral ball valve prosthesis: diagnosis by phonocardiography and echocardiography.

Authors:  I Belenkie; M Carr; R C Schlant; D O Nutter; P N Symbas
Journal:  Am Heart J       Date:  1973-09       Impact factor: 4.749

6.  Alternation of the mitral opening click time in a patient with the Beall mitral valve prosthesis.

Authors:  A Benchimol; C L Harris; K B Desser
Journal:  Chest       Date:  1973-09       Impact factor: 9.410

7.  Phonocardiographic and echocardiographic features of Lillehei-Kaster mitral prosthesis.

Authors:  R Estevez; S Mookherjee; J Potts; M Fulton; A I Obeid
Journal:  J Clin Ultrasound       Date:  1977-06       Impact factor: 0.910

8.  Noninvasive diagnostic evaluation of the normal Beall mitral prosthesis.

Authors:  R A Smith; R E Kerber; J W Snyder
Journal:  Cathet Cardiovasc Diagn       Date:  1976

9.  Mitral valve closure in atrial flutter.

Authors:  M A Greenberg; L S Herman; M V Cohen
Journal:  Circulation       Date:  1979-05       Impact factor: 29.690

10.  CREATION OF A SECONDARY VENTRICULAR OUTLET TO REVERSE HEMOLYSIS AFTER AORTIC VALVE REPLACEMENT.

Authors:  David A. Ott; Denton A. Cooley; John C. Norman
Journal:  Cardiovasc Dis       Date:  1979-09
View more
  1 in total

1.  Noninvasive assessment of the Ionescu-Shiley pericardial xenograft in the mitral position: preliminary experience.

Authors:  L W Pechacek; L G Solana; C M DeCastro; S K Edelman; E Garcia; R J Hall
Journal:  Tex Heart Inst J       Date:  1982-09