Abstract

The objective of this study was to provide calibration curves for correcting intraocular pressure (IOP) measurements obtained using the Tono-Pen XL tonometer in cats, cows and sheep. Twelve eyes from 9 cats, 13 eyes from 7 cows, 10 eyes from 5 sheep were used. The anterior chamber of the eye was cannulated in vivo, in situ (immediately post mortem) or ex vivo with a fine needle and IOP was varied from 10 to 90 mmHg in steps of 10 mmHg by adjusting the height of a saline reservoir connected to the needle. For each pressure setting, several readings of IOP were made using the tonometer. The relationship between Tono-Pen reading and manometer setting was linear over the full range of measurement. However, the slope of the data regression line deviated significantly from 1 and indicated that the instrument systematically underestimated IOP. For cats the average slope was 0.62 and for cows and sheep it was 0.72 and 0.69, respectively. For the latter animals, the regression line also had a nonzero intercept of approximately 4.5 mmHg. Similar results were obtained from in vivo and ex vivo eyes and with different Tono-Pen XL tonometers. Although developed for use on humans, the Tono-Pen XL can provide reproducible and accurate measurement of IOP in cats, cows and sheep when suitably calibrated by manometry. The calibration curves provided here, and by implication those reported for other animals using this tonometer, differ in slope from those measured with earlier models of the Tono-Pen. The reproducibility of the curves we obtained implies that they can be used to correct IOP readings from the Tono-Pen XL when manometry is not possible.

Entities:  

Mesh:

Year:  2004        PMID: 15200622      PMCID: PMC6746305          DOI: 10.1111/j.1463-5224.2004.04038.x

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  8 in total

Review 1.  Feline glaucoma--a comprehensive review.

Authors:  Gillian J McLellan; Paul E Miller
Journal:  Vet Ophthalmol       Date:  2011-09       Impact factor: 1.644

2.  The effect of dorzolamide 2% on circadian intraocular pressure in cats with primary congenital glaucoma.

Authors:  Kelly J Sigle; Gabriel Camaño-Garcia; Alicia L Carriquiry; Daniel M Betts; Markus H Kuehn; Gillian J McLellan
Journal:  Vet Ophthalmol       Date:  2011-09       Impact factor: 1.644

3.  Intraocular pressure measurements in cattle, sheep, and goats with 2 different types of tonometers.

Authors:  Nina Peche; Johanna Corinna Eule
Journal:  Can J Vet Res       Date:  2018-07       Impact factor: 1.310

4.  Evaluation of the Perkins handheld applanation tonometer in horses and cattle.

Authors:  Silvia Franco Andrade; Daniel Silva Kupper; Luiz Fernando Rodrigues Pinho; Elizabeth Cunha Franco; Marcus Vinícius Felix Fabri Prataviera; Rodrigo Rolim Duarte; José Ricardo Cecílio Junqueira
Journal:  J Vet Sci       Date:  2011-06       Impact factor: 1.672

5.  Validation of the TonoVet® rebound tonometer in normal and glaucomatous cats.

Authors:  Gillian J McLellan; Jeremy P Kemmerling; Julie A Kiland
Journal:  Vet Ophthalmol       Date:  2012-06-06       Impact factor: 1.644

6.  Evaluation of tensile strength of tissue adhesives and sutures for clear corneal incisions using porcine and bovine eyes, with a novel standardized testing platform.

Authors:  Simon Kaja; Daryl L Goad; Fatima Ali; Ashley Abraham; R Luke Rebenitsch; Savak Teymoorian; Rohit Krishna; Peter Koulen
Journal:  Clin Ophthalmol       Date:  2012-02-28

7.  Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs.

Authors:  João Victor Goulart Consoni Passareli; Felipe Franco Nascimento; Giovana José Garcia Estanho; Claudia Lizandra Ricci; Glaucia Prada Kanashiro; Rogério Giuffrida; Silvia Franco Andrade
Journal:  Vet World       Date:  2021-09-21

8.  Management of bilateral uveitis in a Toxoplasma gondii-seropositive cat with histopathologic evidence of fungal panuveitis.

Authors:  J Pearce; E A Giuliano; L E Galle; G Klauss; J Ota; C P Moore
Journal:  Vet Ophthalmol       Date:  2007 Jul-Aug       Impact factor: 1.644

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.