Lag measurement in an a-Se active matrix flat-panel imager.

C Schroeder1, T Stanescu, S Rathee, B G Fallone.   

Abstract

Lag and residual contrast have been quantified in an amorphous selenium (a-Se) active-matrix flat-panel imager (AMFPI) as a function of frame time, kilovoltage (kV) and megavoltage (MV) x-ray photon energies and amount of radiation incident on the detector. The AMFPI contains a 200 microm thick a-Se layer deposited on a thin film transistor (TFT) array of size 8.7 cm x 8.7 cm with an 85-microm pixel pitch. For all energies, the lag (signal normalized to the signal due to exposure) for the first (n = 1) and second (n = 2) frame after exposure ranges from 0.45% to 0.91% and from 0.29% to 0.51%, respectively. The amount of lag was determined to be a function of the time after the x-ray exposure irrespective of frame time or the magnitude of exposure. The lag for MV photon energies was slightly less than that for kV photon energies. The residual contrast for all energies studied ranges from 0.41% to 0.75% and from 0.219% to 0.41% for the n = 1 and n = 2 frames, respectively. These results show that lag and residual contrast in kV and MV radiographic applications are always less than 1% for the detection system used and only depend on the time after x-ray exposure.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15191310     DOI: 10.1118/1.1713298

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Simulation system for understanding the lag effect in fluoroscopic images.

Authors:  Rie Tanaka; Hiroki Kawashima; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2012-12-29

2.  Investigation of image lag and modulation transfer function in fluoroscopy images obtained with a dynamic flat-panel detector.

Authors:  Hiroki Kawashima; Rie Tanaka; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Iida; Shigeru Sanada
Journal:  Radiol Phys Technol       Date:  2013-04-09

3.  Evaluation of image lag in a flat-panel, detector-equipped cardiovascular X-ray machine using a newly developed dynamic phantom.

Authors:  Hisaya Sato; Keisuke Kondo; Kyoichi Kato; Yasuo Nakazawa
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.