| Literature DB >> 31061601 |
Tankeshwar Boruah1, Sapan Kumar1, Mohit Kumar Patralekh1, Shambhu Prashad1, Vibash Chandra1, Ijack Debbarma1, Ramesh Kumar1.
Abstract
Reaming is an important step in long bone nailing and has a low complication rate. We report a case of a flexible reamer that got broken and incarcerated in the femoral canal during reaming in a segmental femur fracture. Routine extraction using guide wire was not possible. The incarcerated reamer was successfully extracted with the help of a jumbo cutter.Entities:
Keywords: Femur; Fracture fixation; Intramedullary
Year: 2018 PMID: 31061601 PMCID: PMC6492215 DOI: 10.1016/j.jcot.2018.09.003
Source DB: PubMed Journal: J Clin Orthop Trauma ISSN: 0976-5662