Literature DB >> 15116403

Effect of slide track shape on the wear of ultra-high molecular weight polyethylene in a pin-on-disk wear simulation of total hip prosthesis.

Vesa Saikko1, Olof Calonius, Jaakko Keränen.   

Abstract

Prosthetic joints appear to show a strong relationship between the type of relative motion and wear, requiring careful consideration in the design of wear simulators. This relationship was studied with a 12-station pin-on-disk device, specifically adapted for the wear simulation of prosthetic hip joints. Each station had a unique motion, characterized by the so-called slide track, the track of the pin on the disk. The slide track shapes included 10 ellipses, their aspect ratio (AR) varying from 1.1 to 11.0, and a circle and a straight line as extreme cases. Hence for the first time in hip wear simulation, the motion was systematically varied over a wide range. Conventional UHMWPE pins were tested against polished CoCr disks in diluted calf serum three times for 3 million cycles. Below the AR value of 5.5, the polyethylene wear factor and wear mechanisms agreed with clinical observations. Above this value, the wear factor decreased to unrealistically low values, and the wear surface topography differed from that of retrieved acetabular cups. The wear particles, however, were similar to those isolated from periprosthetic tissues, irrespective of the AR value. In conclusion, it is recommended that the AR value be kept well below the critical point of 5.5. Copyright 2004 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15116403     DOI: 10.1002/jbm.b.20043

Source DB:  PubMed          Journal:  J Biomed Mater Res B Appl Biomater        ISSN: 1552-4973            Impact factor:   3.368


  5 in total

1.  Quantification of the effect of cross-shear on the wear of conventional and highly cross-linked UHMWPE.

Authors:  Lu Kang; Alison L Galvin; Thomas D Brown; Zhongmin Jin; John Fisher
Journal:  J Biomech       Date:  2007-10-22       Impact factor: 2.712

2.  Synthesis and Mechanical Wear Studies of Ultra Smooth Nanostructured Diamond (USND) Coatings Deposited by Microwave Plasma Chemical Vapor Deposition with He/H(2)/CH(4)/N(2) Mixtures.

Authors:  S Chowdhury; J Borham; S A Catledge; A W Eberhardt; P S Johnson; Y K Vohra
Journal:  Diam Relat Mater       Date:  2008       Impact factor: 3.315

Review 3.  Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers.

Authors:  D Baykal; R S Siskey; H Haider; V Saikko; T Ahlroos; S M Kurtz
Journal:  J Mech Behav Biomed Mater       Date:  2013-05-29

4.  Preliminary tribological evaluation of nanostructured diamond coatings against ultra-high molecular weight polyethylene.

Authors:  Michael R Hill; Shane A Catledge; Valeriy Konovalov; William C Clem; Shafiul A Chowdhury; Brandon S Etheridge; Andrei Stanishevsky; Jack E Lemons; Yogesh K Vohra; Alan W Eberhardt
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2008-04       Impact factor: 3.368

Review 5.  Effects of surface coating on reducing friction and wear of orthopaedic implants.

Authors:  Hee Ay Ching; Dipankar Choudhury; Md Julker Nine; Noor Azuan Abu Osman
Journal:  Sci Technol Adv Mater       Date:  2014-01-07       Impact factor: 8.090

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.