Literature DB >> 15077658

Durability of implanted electrodes and leads in an upper-limb neuroprosthesis.

Kevin L Kilgore1, P Hunter Peckham, Michael W Keith, Fred W Montague, Ronald L Hart, Martha M Gazdik, Anne M Bryden, Scott A Snyder, Thomas G Stage.   

Abstract

Implanted neuroprosthetic systems have been successfully used to provide upper-limb function for over 16 years. A critical aspect of these implanted systems is the safety, stability, and-reliability of the stimulating electrodes and leads. These components are (1) the stimulating electrode itself, (2) the electrode lead, and (3) the lead-to-device connector. A failure in any of these components causes the direct loss of the capability to activate a muscle consistently, usually resulting in a decrement in the function provided by the neuroprosthesis. Our results indicate that the electrode, lead, and connector system are extremely durable. We analyzed 238 electrodes that have been implanted as part of an upper-limb neuroprosthesis. Each electrode had been implanted at least 3 years, with a maximum implantation time of over 16 years. Only three electrode-lead failures and one electrode infection occurred, for a survival rate of almost 99 percent. Electrode threshold measurements indicate that the electrode response is stable over time, with no evidence of electrode migration or continual encapsulation in any of the electrodes studied. These results have an impact on the design of implantable neuroprosthetic systems. The electrode-lead component of these systems should no longer be considered a weak technological link.

Entities:  

Mesh:

Year:  2003        PMID: 15077658     DOI: 10.1682/jrrd.2003.11.0457

Source DB:  PubMed          Journal:  J Rehabil Res Dev        ISSN: 0748-7711


  20 in total

1.  An implanted upper-extremity neuroprosthesis using myoelectric control.

Authors:  Kevin L Kilgore; Harry A Hoyen; Anne M Bryden; Ronald L Hart; Michael W Keith; P Hunter Peckham
Journal:  J Hand Surg Am       Date:  2008-04       Impact factor: 2.230

2.  Implanted neuroprosthesis for restoring arm and hand function in people with high level tetraplegia.

Authors:  William D Memberg; Katharine H Polasek; Ronald L Hart; Anne M Bryden; Kevin L Kilgore; Gregory A Nemunaitis; Harry A Hoyen; Michael W Keith; Robert F Kirsch
Journal:  Arch Phys Med Rehabil       Date:  2014-02-20       Impact factor: 3.966

3.  Mechanical fatigue resistance of an implantable branched lead system for a distributed set of longitudinal intrafascicular electrodes.

Authors:  A E Pena; S S Kuntaegowdanahalli; J J Abbas; J Patrick; K W Horch; R Jung
Journal:  J Neural Eng       Date:  2017-12       Impact factor: 5.379

4.  Characterization of high capacitance electrodes for the application of direct current electrical nerve block.

Authors:  Tina Vrabec; Niloy Bhadra; Jesse Wainright; Narendra Bhadra; Manfred Franke; Kevin Kilgore
Journal:  Med Biol Eng Comput       Date:  2015-09-11       Impact factor: 2.602

5.  Selective stimulation of the human femoral nerve with a flat interface nerve electrode.

Authors:  M A Schiefer; K H Polasek; R J Triolo; G C J Pinault; D J Tyler
Journal:  J Neural Eng       Date:  2010-03-08       Impact factor: 5.379

6.  Tension and fatigue behavior of 316LVM 1x7 multi-strand cables used as implantable electrodes.

Authors:  John J Lewandowski; Ravikumar Varadarajan; Brian Smith; Chris Tuma; Mostafa Shazly; Luciano O Vatamanu
Journal:  Mater Sci Eng A Struct Mater       Date:  2008-07-15       Impact factor: 5.234

7.  Challenges and opportunities in restoring function after paralysis.

Authors:  P Hunter Peckham; Kevin L Kilgore
Journal:  IEEE Trans Biomed Eng       Date:  2013-03-07       Impact factor: 4.538

Review 8.  Technological advances in interventions to enhance poststroke gait.

Authors:  Lynne R Sheffler; John Chae
Journal:  Phys Med Rehabil Clin N Am       Date:  2013-05       Impact factor: 1.784

9.  Automated optimal coordination of multiple-DOF neuromuscular actions in feedforward neuroprostheses.

Authors:  J Luis Lujan; Patrick E Crago
Journal:  IEEE Trans Biomed Eng       Date:  2009-01       Impact factor: 4.538

10.  Evolution of Neuroprosthetic Approaches to Restoration of Upper Extremity Function in Spinal Cord Injury.

Authors:  Kevin L Kilgore; Anne Bryden; Michael W Keith; Harry A Hoyen; Ronald L Hart; Gregory A Nemunaitis; P Hunter Peckham
Journal:  Top Spinal Cord Inj Rehabil       Date:  2018
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.