Time dependent variations in biomechanical properties of cadaveric fascia, porcine dermis, porcine small intestine submucosa, polypropylene mesh and autologous fascia in the rabbit model: implications for sling surgery.

Chandler D Dora1, David S Dimarco, Mark E Zobitz, Daniel S Elliott.   

Abstract

PURPOSE: We investigated time dependent variations in tensile strength, stiffness, shrinkage and distortion in 6 materials commonly used for transvaginal anti-incontinence surgery.
MATERIALS AND METHODS: A total of 15 rabbits were randomized into 3 survival groups (2, 6 and 12 weeks, respectively). Each rabbit had human cadaveric fascia, porcine dermis, porcine small intestine submucosa, polypropylene mesh and autologous fascia implanted on the anterior rectus fascia. At harvest tensiometry and image analysis were performed on each sling. Results were compared to baseline for each sling type and the percent decrease from baseline was compared among sling types.
RESULTS: Each type of human cadaveric fascia and porcine allografts showed a marked decrease (60% to 89%) in tensile strength and stiffness from baseline. Polypropylene mesh and autologous fascia did not differ in tensile strength from baseline. Polypropylene mesh increased in stiffness from baseline. Autologous fascia and small intestinal submucosa demonstrated a 41% and 50% decrease in surface area, respectively, at 12 weeks.
CONCLUSIONS: To our knowledge the relative contribution of biomechanical properties of sling material to the success of anti-incontinence surgery is unknown. However, rapid loss of tensile strength and stiffness in porcine and cadaveric materials may contribute to the early re-emergence of symptoms following successful sling surgery. The results of this study add scientific validity to the increasing use of synthetics in anti-incontinence surgery. Urologists selecting a sling material should be aware of its time dependent biomechanical fate relative to other sling materials.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15076323     DOI: 10.1097/01.ju.0000121377.61788.ad

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450


  21 in total

Review 1.  Evaluation of current biologic meshes in pelvic organ prolapse repair.

Authors:  Ashley Cox; Sender Herschorn
Journal:  Curr Urol Rep       Date:  2012-06       Impact factor: 3.092

Review 2.  InteXen tissue processing and laboratory study.

Authors:  J Christian Winters
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2006-06

3.  Cadaveric fascia lata.

Authors:  Pamela A Moalli
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2006-06

4.  The effect of source animal age upon extracellular matrix scaffold properties.

Authors:  Stephen Tottey; Scott A Johnson; Peter M Crapo; Janet E Reing; Li Zhang; Hongbin Jiang; Christopher J Medberry; Brandon Reines; Stephen F Badylak
Journal:  Biomaterials       Date:  2010-09-25       Impact factor: 12.479

5.  Correlation between vaginal stiffness index and pelvic floor disorder quality-of-life scales.

Authors:  Lee B Epstein; Carol A Graham; Michael H Heit
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2008-01-24

6.  Urinary bladder matrix promotes site appropriate tissue formation following right ventricle outflow tract repair.

Authors:  Nathaniel T Remlinger; Thomas W Gilbert; Masahiro Yoshida; Brogan N Guest; Ryotaro Hashizume; Michelle L Weaver; William R Wagner; Bryan N Brown; Kimimasa Tobita; Peter D Wearden
Journal:  Organogenesis       Date:  2013-06-25       Impact factor: 2.500

Review 7.  Polypropylene mesh and the host response.

Authors:  Hiren Patel; Donald R Ostergard; Gina Sternschuss
Journal:  Int Urogynecol J       Date:  2012-03-20       Impact factor: 2.894

8.  Tensile properties of commonly used prolapse meshes.

Authors:  Keisha A Jones; Andrew Feola; Leslie Meyn; Steven D Abramowitch; Pamela A Moalli
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2009-03-28

9.  Investigation of soft-tissue stiffness alteration in denervated human tissue using an ultrasound indentation system.

Authors:  Mohsen Makhsous; Ganapriya Venkatasubramanian; Aditya Chawla; Yagna Pathak; Michael Priebe; William Z Rymer; Fang Lin
Journal:  J Spinal Cord Med       Date:  2008       Impact factor: 1.985

10.  Evaluation of xenogenic extracellular matrices as adjuvant scaffolds for the treatment of stress urinary incontinence.

Authors:  Niall F Davis; Daniel N Coakley; Anthony Callanan; Hugh D Flood; Tim M McGloughlin
Journal:  Int Urogynecol J       Date:  2013-06-29       Impact factor: 2.894

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.