Comparison of complementary metal oxide semiconductor and charge-coupled device intraoral X-ray detectors using subjective image quality.

H Kitagawa1, J P Scheetz, A G Farman.   

Abstract

OBJECTIVE: To compare the subjective image quality of the newer generation Schick CDR detector employing complementary metal oxide semiconductor (CMOS) technology with images using the earlier generation charge-coupled device (CCD) Schick CDR detector.
METHODS: All radiographic images were made using the same formalin-fixed adult cadaver maxilla with surrounding natural soft tissues in place. The X-ray generator used was a Villa Sistemi Medicali Diamatic srl AP/Explor X operated at 70 kVp and 8 mA. The source-to-detector distance was set at 38 cm and an optical bench was used to ensure reproducible beam geometry. A range of exposures was applied for both detectors. A panel of nine dentists independently observed and evaluated images made at each exposure. For both detectors, the three images ranked highest were randomized for re-evaluation in panels of six images. Each image was repeated randomly a total of 10 times. Features chosen as observation points were: (1) proximal dental caries; (2) gingival soft tissues; (3) cortical bone; (4) root canal space; (5) root apices; (6) periodontal ligament space; and (7) endodontic instrument tip clarity. Comparisons were made by use of odds ratio analysis applying a 95% confidence level. Interrater and intrarater reliabilities were computed to assess consistency in observer ratings.
RESULTS: The CMOS sensor was rated as outperforming its CCD predecessor for depiction of cortical bone and root apices; the CCD detector was only rated superior for depiction of root canal space. No significant difference was found between the two detectors in perceived depiction of proximal dental caries, gingival soft tissues, periodontal ligament space or endodontic instruments. Combining rating scores from each of the tasks, CMOS and CCD detectors had a similar proportion of image ratings of excellent, acceptable and poor.
CONCLUSIONS: Regarding subjective image quality, the Schick CMOS and CCD detectors were perceived to produce radiographic images of similar overall quality.

Entities:  

Mesh:

Year:  2003        PMID: 15070845     DOI: 10.1259/dmfr/19990417

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  9 in total

1.  Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study.

Authors:  K Hellén-Halme; A Lith
Journal:  Dentomaxillofac Radiol       Date:  2012-03       Impact factor: 2.419

Review 2.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

3.  Diagnostic accuracy of different imaging modalities in detection of proximal caries.

Authors:  B Senel; K Kamburoglu; O Uçok; S P Yüksel; T Ozen; H Avsever
Journal:  Dentomaxillofac Radiol       Date:  2010-12       Impact factor: 2.419

4.  A Comparison of intra-oral digital imaging modalities: Charged Couple Device versus Storage Phosphor Plate.

Authors:  A Anas; Jm Asaad; Ka Tarboush
Journal:  Int J Health Sci (Qassim)       Date:  2010-11

5.  The Effects on Absorbed Dose Distribution in Intraoral X-ray Imaging When Using Tube Voltages of 60 and 70 kV for Bitewing Imaging.

Authors:  Kristina Hellén-Halme; Mats Nilsson
Journal:  J Oral Maxillofac Res       Date:  2013-10-01

6.  A comparative study on image quality of two digital intraoral sensors.

Authors:  Cinar Aziman; Kristina Hellén-Halme; Xie-Qi Shi
Journal:  Dentomaxillofac Radiol       Date:  2019-05-14       Impact factor: 2.419

7.  Diagnostic Accuracy of Charge-coupled Device Sensor and Photostimulable Phosphor Plate Receptor in the Detection of External Root Resorption In Vitro.

Authors:  Shirin Sakhdari; Zohreh Khalilak; Elham Najafi; Rezvaneh Cheraghi
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2015-03-04

8.  The effects of image acquisition control of digital X-ray system on radiodensity quantification.

Authors:  Wook-Jin Seong; Hyeon-Cheol Kim; Soocheol Jeong; Youngcheul Heo; Woo-Bin Song; Mansur Ahmad
Journal:  Restor Dent Endod       Date:  2013-08-23

9.  Characteristics of radiographic images acquired with CdTe, CCD and CMOS detectors in skull radiography.

Authors:  Polyane Mazucatto Queiroz; Gustavo Machado Santaella; Sergio Lucio Pereira de Castro Lopes; Francisco Haiter-Neto; Deborah Queiroz Freitas
Journal:  Imaging Sci Dent       Date:  2020-12-15
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.