Effect of protein on hemoglobin and hematocrit assays with a conductivity-based point-of-care testing device: comparison with optical methods.

Sidney M Hopfer1, Francesca L Nadeau, Marilyn Sundra, Gregory S Makowski.   

Abstract

Point-of-care testing (POCT) for blood hemoglobin and hematocrit (H/H) levels provides rapid patient assessment including the need for transfusion. Conductivity-based methods of blood H/H determinations can be influenced by plasma protein concentration. To assess this factor, we measured H/H levels at varying protein concentrations using two POCT instruments: iSTAT-1 (conductivity method) and Hemocue (optical method). These H/H results were compared to results obtained by our core laboratory hematology analyzer (GenS). Anticoagulated whole blood was centrifuged to sediment the red blood cells; the plasma was removed to serve as source of protein for mixing studies. A series of reconstituted samples was prepared with varying H/H and protein levels. To mimic hemodilution in clinical practice, samples were diluted with saline or lactated Ringer's solutions. Following H/H analysis, the samples were centrifuged and protein determined in the supernatant plasmas. The H/H results obtained with the Hemocue instrument correlated exactly with those of the GenS analyzer at protein concentrations from 0.7 to 6.2 g/dl. The correlation was unaffected when hemodilution was performed with either saline (r = 0.999) or lactated Ringer's (r = 1.000). The H/H results obtained with the iSTAT-1 instrument gave slightly less correlation with those of the GenS analyzer (r = 0.978 - 0.980) over this protein range. However, the iSTAT-1 results were generally lower than the GenS results, with discrepancies up to 2 g/dL for hemoglobin values and up to 4% for hematocrits at the lowest protein concentration. Therefore, it is recommended that H/H testing in patients with suspected hypoproteinemia or substantial hemodilution should be tested with a non-conductivity-based method.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15038671

Source DB:  PubMed          Journal:  Ann Clin Lab Sci        ISSN: 0091-7370            Impact factor:   1.256


  10 in total

1.  Conductivity-based hematocrit measurement during cardiopulmonary bypass.

Authors:  Jacoline Steinfelder-Visscher; Patrick W Weerwind; Steven Teerenstra; Gheorghe A M Pop; René M H J Brouwer
Journal:  J Clin Monit Comput       Date:  2006-11-04       Impact factor: 2.502

2.  Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery.

Authors:  Kirti P Patel; Gary W Hay; Mahesh Keitheri Cheteri; David W Holt
Journal:  J Extra Corpor Technol       Date:  2007-03

Review 3.  Factors affecting hemoglobin measurement.

Authors:  Lauren Berkow
Journal:  J Clin Monit Comput       Date:  2013-03-26       Impact factor: 2.502

4.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

5.  Clinical evaluation of the ABL-77 for point-of-care analysis in the cardiovascular operating room.

Authors:  Jack S Prichard; John S French; Nestor Alvar
Journal:  J Extra Corpor Technol       Date:  2006-06

6.  Venous blood gases, plasma biochemistry, and hematology of wild-caught common chameleons (Chamaeleo chamaeleon).

Authors:  David Eshar; Melanie Ammersbach; Boaz Shacham; Gad Katzir; Hugues Beaufrère
Journal:  Can J Vet Res       Date:  2018-04       Impact factor: 1.310

7.  Performance of point-of-care diagnostics for glucose, lactate, and hemoglobin in the management of severe malaria in a resource-constrained hospital in Uganda.

Authors:  Michael Hawkes; Andrea L Conroy; Robert O Opoka; Sophie Namasopo; W Conrad Liles; Chandy C John; Kevin C Kain
Journal:  Am J Trop Med Hyg       Date:  2014-03-03       Impact factor: 2.345

8.  Comparison of four different hematocrit assays and the effect of albumin on their measurements.

Authors:  Amelie Pare; Laura Kippen; Catherine Wagg; Matt Longmore; Soren Boysen
Journal:  Front Vet Sci       Date:  2022-08-24

9.  Blood parameters as biomarkers in a Salmonella spp. disease model of weaning piglets.

Authors:  Emili Barba-Vidal; Victor Fernando Buttow Roll; Edgar Garcia Manzanilla; Carlos Torrente; José Antonio Moreno Muñoz; José Francisco Pérez; Susana María Martín-Orúe
Journal:  PLoS One       Date:  2017-10-26       Impact factor: 3.240

10.  Influence of measurement principle on total hemoglobin value.

Authors:  Keisuke Hayashi; Takashi Hitosugi; Yoshifumi Kawakubo; Norihisa Kitamoto; Takeshi Yokoyama
Journal:  BMC Anesthesiol       Date:  2020-04-07       Impact factor: 2.217

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.