Literature DB >> 15028076

Implications of implantable cardioverter defibrillator therapy in congenital heart disease and pediatrics.

Mark E Alexander1, Frank Cecchin, Edward P Walsh, John K Triedman, Laura M Bevilacqua, Charles I Berul.   

Abstract

INTRODUCTION: The incidence of appropriate and inappropriate discharges, indicators of system failure, and clinical implications of implantable cardioverter defibrillator (ICD) therapy in children and young adults with heart disease is poorly defined. METHODS AND
RESULTS: In a retrospective study at a single medical center, a total of 90 ICD procedures were performed in 76 patients younger than age 30 years (median 16 years, range 1-30): 42% with congenital heart disease, 33% with primary electrical disease, 17% with hypertrophic cardiomyopathy, and 8% with idiopathic dilated cardiomyopathy. Indications for ICD included arrest or sustained ventricular tachycardia (n = 27), and combinations of syncope (n = 32), palpitations (n = 17), spontaneous ventricular arrhythmia (n = 40), inducible ventricular tachycardia (n = 36), or severe hypertrophic cardiomyopathy. Transvenous dual-chamber ICDs were implanted in 29 patients. Subcutaneous arrays or epicardial patches were used in 9 patients. Over a median 2-year follow-up, 28% of patients received appropriate shocks for ventricular tachycardia (median 13 months to first shock) and 25% experienced inappropriate shocks for multiple causes (median 16 months). With multivariate analysis, growth strongly correlated with lead failure (odds ratio 73, 3.5-1530, P = 0.006). Complications occurred in 29 patients, including lead failure in 16 (21%), ICD "storm" with sequential shocks in 5, and infection in 2 patients. No deaths were attributable to ICD placement or subsequent device failure.
CONCLUSION: ICD therapy can effectively manage malignant arrhythmias in selected pediatric and congenital heart patients. Spurious shocks or ICD storm may increase morbidity and emphasize the need for concomitant medical and ablative therapy. ICD lead failure was relatively frequent in this population.

Entities:  

Mesh:

Year:  2004        PMID: 15028076     DOI: 10.1046/j.1540-8167.2004.03388.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  25 in total

Review 1.  Hypertrophic cardiomyopathy in childhood.

Authors:  Steven D Colan
Journal:  Heart Fail Clin       Date:  2010-10       Impact factor: 3.179

2.  Validating defibrillation simulation in a human-shaped phantom.

Authors:  Jess D Tate; Thomas A Pilcher; Kedar K Aras; Brett M Burton; Rob S MacLeod
Journal:  Heart Rhythm       Date:  2019-11-23       Impact factor: 6.343

3.  Measuring defibrillator surface potentials: The validation of a predictive defibrillation computer model.

Authors:  Jess Tate; Jeroen Stinstra; Thomas Pilcher; Ahrash Poursaid; Matthew A Jolley; Elizabeth Saarel; John Triedman; Rob S MacLeod
Journal:  Comput Biol Med       Date:  2018-08-29       Impact factor: 4.589

4.  Long-term follow-up of children and adolescents diagnosed with hypertrophic cardiomyopathy: risk factors for adverse arrhythmic events.

Authors:  Jeffrey P Moak; Eric S Leifer; Dorothy Tripodi; Saidi A Mohiddin; Lameh Fananapazir
Journal:  Pediatr Cardiol       Date:  2011-04-13       Impact factor: 1.655

5.  Prospective evaluation of defibrillation threshold and postshock rhythm in young ICD recipients.

Authors:  Andrew E Radbill; John K Triedman; Charles I Berul; Edward P Walsh; Mark E Alexander; Gregory Webster; Frank Cecchin
Journal:  Pacing Clin Electrophysiol       Date:  2012-09-14       Impact factor: 1.976

6.  Placement of implantable cardioverter-defibrillators in paediatric and congenital heart defect patients: a pipeline for model generation and simulation prediction of optimal configurations.

Authors:  Lukas J Rantner; Fijoy Vadakkumpadan; Philip J Spevak; Jane E Crosson; Natalia A Trayanova
Journal:  J Physiol       Date:  2013-06-24       Impact factor: 5.182

Review 7.  Surgical techniques for implanting implantable cardioverter defibrillators in children and infants.

Authors:  Shoji Suzuki; Shinya Motohashi; Masahiko Matsumoto
Journal:  Surg Today       Date:  2013-10-24       Impact factor: 2.549

8.  Congenital Heart Defects in Adults : A Field Guide for Cardiologists.

Authors:  Anitra Romfh; Francesca Romana Pluchinotta; Prashob Porayette; Anne Marie Valente; Stephen P Sanders
Journal:  J Clin Exp Cardiolog       Date:  2012-06-15

9.  A computer modeling tool for comparing novel ICD electrode orientations in children and adults.

Authors:  Matthew Jolley; Jeroen Stinstra; Steve Pieper; Rob Macleod; Dana H Brooks; Frank Cecchin; John K Triedman
Journal:  Heart Rhythm       Date:  2008-01-17       Impact factor: 6.343

10.  Update on the use and outcomes of implantable cardioverter defibrillators in pediatric patients.

Authors:  Christian J Turner; Elizabeth A Stephenson
Journal:  Curr Treat Options Cardiovasc Med       Date:  2012-10
View more