Surface analysis of early retrieved acetabular polyethylene liners: a comparison of conventional and highly crosslinked polyethylenes.

Orhun K Muratoglu1, Evan S Greenbaum, Charles R Bragdon, Murali Jasty, Andrew A Freiberg, William H Harris.   

Abstract

In vivo wear behavior of 16 highly cross-linked and 19 conventional polyethylene acetabular explants, at an average in vivo duration of 6 months were studied. Highly cross-linked groups showed machining marks from the original manufacturing process in some areas and extensive scratching and some polishing of the articulating surfaces. The conventional group showed greater loss of machining marks, scratching, and polishing. Representative samples were melted to allow recovery of plastic deformation and to show true removal of material caused by wear. Melt-recovery experiments consistently showed the disappearance of surface scratches and the restoration of the original machining marks in the highly cross-linked explants. In the conventional group, few of the scratches were eliminated and only limited restoration of the machining marks was apparent. These observations support the hypothesis that the early in vivo scratching of highly cross-linked polyethylene acetabular liners is primarily caused by plastic deformation.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14716654     DOI: 10.1016/j.arth.2003.08.003

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  14 in total

1.  Mechanical properties of retrieved highly cross-linked crossfire liners after short-term implantation.

Authors:  Steven M Kurtz; William Hozack; Joseph Turner; James Purtill; Daniel MacDonald; Peter Sharkey; Javad Parvizi; Michael Manley; Richard Rothman
Journal:  J Arthroplasty       Date:  2005-10       Impact factor: 4.757

2.  Wear is reduced in THA performed with highly cross-linked polyethylene.

Authors:  Burak Beksaç; Antonio Salas; Alejandro González Della Valle; Eduardo A Salvati
Journal:  Clin Orthop Relat Res       Date:  2008-12-12       Impact factor: 4.176

3.  Early failure of a cross-linked polyethylene acetabular liner. A case report.

Authors:  K David Moore; Preston R Beck; Donald W Petersen; John M Cuckler; Jack E Lemons; Alan W Eberhardt
Journal:  J Bone Joint Surg Am       Date:  2008-11       Impact factor: 5.284

4.  Early polyethylene wear and excessive acetabular granuloma in an uncemented HA-coated total hip arthroplasty--midterm results of a prospective study.

Authors:  Jörg Axel Karl Ohnsorge; Jack Davis; Uwe Maus; Muharrem Saklak; Markus Weisskopf; Dieter Christian Wirtz
Journal:  HSS J       Date:  2006-09

5.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

6.  Scratching vulnerability of conventional vs highly cross-linked polyethylene liners because of large embedded third-body particles.

Authors:  Anneliese D Heiner; Alison L Galvin; John Fisher; John J Callaghan; Thomas D Brown
Journal:  J Arthroplasty       Date:  2011-11-23       Impact factor: 4.757

7.  Retrieved highly crosslinked UHMWPE acetabular liners have similar wear damage as conventional UHMWPE.

Authors:  David T Schroder; Natalie H Kelly; Timothy M Wright; Michael L Parks
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

8.  Mechanical properties, oxidation, and clinical performance of retrieved highly cross-linked Crossfire liners after intermediate-term implantation.

Authors:  Steven M Kurtz; Matthew S Austin; Khalid Azzam; Peter F Sharkey; Daniel W MacDonald; Francisco J Medel; William J Hozack
Journal:  J Arthroplasty       Date:  2009-06-10       Impact factor: 4.757

9.  Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro.

Authors:  Ana Laura Puente Reyna; Marcus Jäger; Thilo Floerkemeier; Sven Frecher; Karl-Stefan Delank; Christoph Schilling; Thomas M Grupp
Journal:  Biomed Res Int       Date:  2016-09-19       Impact factor: 3.411

10.  UHMWPE for arthroplasty: past or future?

Authors:  Elena Maria Brach Del Prever; Alessandro Bistolfi; Pierangiola Bracco; Luigi Costa
Journal:  J Orthop Traumatol       Date:  2008-12-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.