Use of the Cygnus GlucoWatch biographer at a diabetes camp.

Laura M Gandrud1, Helen U Paguntalan, M Michelle Van Wyhe, Betsy L Kunselman, Amy D Leptien, Darrell M Wilson, Richard C Eastman, Bruce A Buckingham.   

Abstract

BACKGROUND: Detection and prevention of nocturnal hypoglycemia is a major medical concern at diabetes camps.
OBJECTIVE: We conducted an open-label trial of the Cygnus GlucoWatch biographer to detect nocturnal hypoglycemia in a diabetes camp, a nonclinical environment with multiple activities.
METHODS: Forty-five campers (7-17 years old) wore a biographer. The biographer was placed on the arm at 6:00 PM, with the low alarm set to 85 mg/dL (4.7 mmol/L). Overnight glucose monitoring occurred per usual camp protocol. Counselors were to check and record blood glucose values if the biographer alarmed.
RESULTS: Biographers were worn for 154 nights by 45 campers. After a 3-hour warm-up period, 67% of biographers were calibrated, of which 28% were worn the entire night (12 hours). Thirty-four percent of readings were skipped because of: "data errors" (65%), sweat (20%), and temperature change (16%). Reported biographer values correlated with meter glucose values measured 11 to 20 minutes later (r = 0.90). Of 20 low-glucose alarms with corresponding meter values measured within 20 minutes, there were 10 true-positive alarms, 10 false-positive alarms, and no false-negative alarms. Campers reported sleep disruption 32% of the nights, and 74% found the biographer helpful. Campers reported they would wear the biographer 4 to 5 nights each week.
CONCLUSIONS: Half of the biographer low-glucose alarms that had corresponding blood meter values were true-positive alarms, and the remaining were false-positive alarms. There was close correlation between the biographer and meter glucose values. The majority of campers found the biographer helpful and would use it at home.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14702457     DOI: 10.1542/peds.113.1.108

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  5 in total

1.  Response to nocturnal alarms using a real-time glucose sensor.

Authors:  Bruce Buckingham; Jen Block; Jonathan Burdick; Andrea Kalajian; Craig Kollman; Michael Choy; Darrell M Wilson; Peter Chase
Journal:  Diabetes Technol Ther       Date:  2005-06       Impact factor: 6.118

Review 2.  Hypo- and Hyperglycemic Alarms: Devices and Algorithms.

Authors:  Daniel Howsmon; B Wayne Bequette
Journal:  J Diabetes Sci Technol       Date:  2015-04-30

3.  Carbon nanotube composites for glucose biosensor incorporated with reverse iontophoresis function for noninvasive glucose monitoring.

Authors:  Tai-Ping Sun; Hsiu-Li Shieh; Congo Tak-Shing Ching; Yan-Dong Yao; Su-Hua Huang; Chia-Ming Liu; Wei-Hao Liu; Chung-Yuan Chen
Journal:  Int J Nanomedicine       Date:  2010-05-13

Review 4.  Recent progress in mechanical artificial pancreas.

Authors:  Masami Hoshino; Yoshikura Haraguchi; Iwanori Mizushima; Motohiro Sakai
Journal:  J Artif Organs       Date:  2009-09-19       Impact factor: 1.731

Review 5.  Camp for Youth With Type 1 Diabetes.

Authors:  Kelly Fegan-Bohm; Jill Weissberg-Benchell; Daniel DeSalvo; Sheila Gunn; Marisa Hilliard
Journal:  Curr Diab Rep       Date:  2016-08       Impact factor: 4.810

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.