Measurements of the relative output factors for CyberKnife collimators.

Cheng Yu1, Gabor Jozsef, Michael L J Apuzzo, Zbigniew Petrovich.   

Abstract

OBJECTIVE: To determine accurately the relative output factors, defined as the ratio of the nominal dose rate for a given collimator to that of the 60-mm collimator. This is particularly important for radiosurgical treatment of functional disorders, such as trigeminal neuralgia, in which a single large radiation dose is delivered to the target with a small collimator, such as the 5-mm collimator for CyberKnife radiosurgery. Numerous studies on the output factors have been reported for the Leksell gamma knife unit but none for the CyberKnife system.
METHODS: Measurements of the relative output factors for all 12 collimators were performed by three different methods: silicon diode, radiographic film, and thermoluminescent dosimetry microcubes. The silicon diode is designed for measurements in small (1-50 MV) photon beams performed in water or air. Film and thermoluminescent dosimetry measurements were performed in a plastic phantom.
RESULTS: The measured relative output factors for the three methods were very similar except for the three smallest collimators (5, 7.5, and 10 mm). The measured difference between the above methods was approximately 2%. The mean value of the output factor for the 5-mm collimator was 0.686 +/- 0.024. The uncertainties of the output factors are expected to increase with the decrease of collimator diameter. They range from approximately 1 to 4% of the relative output factor.
CONCLUSION: The relative output factor can be measured with an acceptable accuracy even for the smallest (5-mm) CyberKnife collimators. This requires the selection of appropriate dosimetric detectors and measuring procedures. The results obtained with the diode are considered more accurate than with the other two methods.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14683553     DOI: 10.1227/01.neu.0000097328.54068.44

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  4 in total

1.  Total scatter factors of small beams: a multidetector and Monte Carlo study.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

Review 2.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

3.  Comparative analysis between 5 mm and 7.5 mm collimators in CyberKnife radiosurgery for trigeminal neuralgia.

Authors:  H Sudahar; P G G Kurup; V Murali; J Velmurugan
Journal:  J Med Phys       Date:  2013-07

4.  Clinical outcomes of CyberKnife stereotactic body radiotherapy for peripheral stage I non-small cell lung cancer.

Authors:  Ze-Tian Shen; Xin-Hu Wu; Bing Li; Xi-Xu Zhu
Journal:  Med Oncol       Date:  2015-02-01       Impact factor: 3.064

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.