Evaluation of an automated, latex-enhanced turbidimetric D-dimer test (advanced D-dimer) and usefulness in the exclusion of acute thromboembolic disease.

David B Wilson1, Kathy M Gard.   

Abstract

Advanced D-dimer (Dade Behring, Marburg, Germany) is a latex-enhanced turbidimetric test for the quantitative determination of D-dimers. To evaluate this assay, examine test performance on 2 instruments (BCS Analyzer and Sysmex CA-1500, Dade Behring), establish negative predictive value cutoffs (NPVCs), and compare the test with the qualitative SimpliRed test (Agen Biomedical, Brisbane, Australia), we evaluated 117 patients for possible acute thromboembolic disease (pulmonary embolus or deep venous thrombosis). Patients selected for the study were those with confirmatory tests (ultrasound, spiral computed tomography, ventilation-perfusion scan, or angiography) and those without certain medical conditions that might be associated with increased D-dimers. An NPVC of 1.8 mg/L for the BCS Analyzer resulted in 41.0% negative results (48/117) and sensitivity, specificity, and negative and positive predictive values of 100%, 48%, 100%, and 21%, respectively. An NPVC of 1.1 mg/L for the Sysmex CA-1500 resulted in 31.6% negative results (37/117) and sensitivity, specificity, and negative and positive predictive values of 100%, 36%, 100%, and 18%, respectively. The SimpliRed had 31.6% negative results (37/117) and sensitivity, specificity, and negative and positive predictive values of 86%, 34%, 96%, and 15%, respectively.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14671982     DOI: 10.1309/U3XC-MGP4-PMM1-AB18

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  4 in total

1.  Pulmonary embolism: investigation of the clinically assessed intermediate risk subgroup.

Authors:  D J Warren; S Matthews
Journal:  Br J Radiol       Date:  2011-09-21       Impact factor: 3.039

2.  Clinical application of the Innovance D-dimer assay in the diagnosis of acute pulmonary thromboembolism.

Authors:  Zhuxiao Bai; Yurong Huang; Chenghua Song; Huimin Liu; Yihui Chen; Haitao Zhang; Xinhong Lu; Yingbo Song; Xin Zhang
Journal:  Exp Ther Med       Date:  2017-04-28       Impact factor: 2.447

Review 3.  Combined use of rapid D-dimer testing and estimation of clinical probability in the diagnosis of deep vein thrombosis: systematic review.

Authors:  Tonya L Fancher; Richard H White; Richard L Kravitz
Journal:  BMJ       Date:  2004-09-21

Review 4.  Cardiopulmonary laboratory biomarkers in the evaluation of acute dyspnea.

Authors:  Natalie R Stokes; Brett W Dietz; Jackson J Liang
Journal:  Open Access Emerg Med       Date:  2016-05-17
  4 in total