Accuracy of measurement of polyethylene wear with use of radiographs of total hip replacements.

Edward Ebramzadeh1, Sophia N Sangiorgio, Federico Lattuada, Joon-Soon Kang, Roberto Chiesa, Harry A McKellop, Lawrence D Dorr.   

Abstract

BACKGROUND: Although a number of methods are used to estimate polyethylene liner wear from radiographs of total hip replacements, there is no consensus with regard to the accuracy of these methods. The purpose of this study was to compare the accuracy of several such measurement methods with use of both laboratory radiographs and routine clinical radiographs.
METHODS: A phantom apparatus was designed to simulate random values of three-dimensional wear, with varying degrees of cup abduction and anteversion, and to obtain anteroposterior and cross-table lateral radiographs with each value. Wear was measured with use of the Charnley duoradiographic method, the Livermore method, and the method described by Dorr and Wan, as well as with use of PolyWare and Hip32 software packages, both with and without three-dimensional measurements. Clinical wear was measured from conventional radiographs made prior to revision surgery in fourteen patients and was compared with wear measured directly from the retrieved liners with use of a coordinate measuring machine.
RESULTS: With laboratory radiographs, median errors were 0.1 mm with the Livermore method and both computerized methods, 0.23 mm with the Charnley method, and 1.7 mm with the method of Dorr and Wan. Maximum errors were between 0.6 mm (Livermore) and 4.3 mm (Dorr and Wan). In contrast, with use of clinical radiographs, median errors ranged between 0.2 mm (Hip32) and 0.6 mm (Dorr and Wan). Maximum errors ranged between 1.8 mm (Dorr and Wan) and 2.5 mm (Livermore).
CONCLUSIONS: With laboratory radiographs, computerized methods of polyethylene wear measurement offered distinctly greater accuracy than did manual methods; however, with clinical radiographs, they offered only slightly better accuracy. Although the increased accuracy of computerized methods may be necessary in research settings, manual methods provided sufficient accuracy for routine clinical assessment of wear.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14668508     DOI: 10.2106/00004623-200312000-00016

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  20 in total

1.  Relationship between the pelvic osteolytic volume on computed tomography and clinical outcome in patients with cementless acetabular components.

Authors:  Ho Hyun Yun; Won Yong Shon; Suk Joo Hong; Jung-Ro Yoon; Jae-Hyuk Yang
Journal:  Int Orthop       Date:  2010-10-07       Impact factor: 3.075

2.  Wear of a 5 megarad cross-linked polyethylene liner: a 6-year RSA study.

Authors:  Stuart A Callary; David G Campbell; Graham Mercer; Kjell G Nilsson; John R Field
Journal:  Clin Orthop Relat Res       Date:  2013-01-19       Impact factor: 4.176

3.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

4.  Is there a risk in placing a ceramic head on a previously implanted trunion?

Authors:  Didier Hannouche; Jérôme Delambre; Frédéric Zadegan; Laurent Sedel; Rémy Nizard
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

5.  Less wear with aluminium-oxide heads than cobalt-chrome heads with ultra high molecular weight cemented polyethylene cups: a ten-year follow-up with radiostereometry.

Authors:  Jon Dahl; Per Söderlund; Bo Nivbrant; Lars Nordsletten; Stephan M Röhrl
Journal:  Int Orthop       Date:  2011-08-26       Impact factor: 3.075

6.  Analysis of polyethylene wear in plain radiographs.

Authors:  Maiken Stilling; Kjeld Søballe; Niels Trolle Andersen; Kristian Larsen; Ole Rahbek
Journal:  Acta Orthop       Date:  2009-12       Impact factor: 3.717

7.  Clinical comparison of polyethylene wear with zirconia or cobalt-chromium femoral heads.

Authors:  Maiken Stilling; Kjeld Anton Nielsen; Kjeld Søballe; Ole Rahbek
Journal:  Clin Orthop Relat Res       Date:  2009-03-27       Impact factor: 4.176

8.  Low wear of a second-generation highly crosslinked polyethylene liner: a 5-year radiostereometric analysis study.

Authors:  Stuart A Callary; John R Field; David G Campbell
Journal:  Clin Orthop Relat Res       Date:  2013-07-27       Impact factor: 4.176

9.  Degeneration of acetabular articular cartilage to bipolar hemiarthroplasty.

Authors:  Kyoung Ho Moon; Jun Soon Kang; Tong Joo Lee; Sang Hyeop Lee; Sung Wook Choi; Man Hee Won
Journal:  Yonsei Med J       Date:  2008-10-31       Impact factor: 2.759

10.  THR Simulator--the software for generating radiographs of THR prosthesis.

Authors:  Tai-Yin Wu; Rong-Sen Yang; Chiou-Shann Fuh; Sheng-Mou Hou; Chen-Kun Liaw
Journal:  BMC Musculoskelet Disord       Date:  2009-01-16       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.