Quantification of the effect of system and object parameters on edge enhancement in phase-contrast radiography.

Edwin F Donnelly1, Ronald R Price, David R Pickens.   

Abstract

The purpose of this study was to evaluate the effects of system parameters (focal spot size, tube voltage, geometry, detector resolution, and image noise) and object characteristics (edge gradient/ shape, composition, thickness, and overlying attenuating material) upon the edge enhancement effect in phase-contrast radiography. Each variable of interest was adjusted and images of a 3 mm lucite phantom were obtained with the other variables remaining constant. A microfocus x-ray source coupled to a CCD camera with an intensifying screen was used to acquire the digital images. Two parameters of image analysis were used to quantify the effects. The edge enhancement index (EEI) was used to measure the absolute degree of edge enhancement, while the edge enhancement to noise ratio (EE/N) was used to measure the conspicuity of the edge enhancement relative to image noise. Little effect on EEI was seen from tube voltage, object thickness, overlying attenuating material, while focal spot size and system geometry demonstrated measurable effects upon the degree of edge enhancement. It was also shown that while the edge enhancement effect over straight edges is highly dependent upon how the edge aligns with the x-ray beam, rounded edges, which better model biological objects, do not suffer from this dependence and the EEI reaches its maximal level at any alignment. Decreasing detector resolution diminished the EEI slightly, but even with pixel sizes of 0.360 x 0.360 mm edge enhancement effects were readily visible. The effect of image noise on EE/N was evaluated using different exposure times showing an expected improvement with longer exposure time with EE/N approaching a plateau at 5 min. Many of the parameters that will go into the design of a future PC-R imaging system have been quantified in terms of their effect on the degree of edge enhancement in the acquired image. These results, taken together, indicate that either a specimen or even clinical breast imaging system could be created with currently available technology. The major limitation to a clinical system would be the low x-ray flux from the microfocal x-ray source.

Entities:  

Mesh:

Year:  2003        PMID: 14655935     DOI: 10.1118/1.1617430

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  12 in total

1.  Development of ultrafast laser-based x-ray in-vivo phase-contrast micro-CT beamline for biomedical applications at Advanced Laser Light Source (ALLS).

Authors:  Russell Kincaid; Andrzej Krol; Sylvain Fourmaux; Jean-Claude Kieffer; Cristina Serbanescu; Marina Servol; Levon Vogelsang; Steve Wilkins; Andrew Stevenson; Yakov Nesterets; Edward Lipson; Hongwei Ye; Andrew Pogany
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008-01-01

2.  Characterization of a high-energy in-line phase contrast tomosynthesis prototype.

Authors:  Di Wu; Aimin Yan; Yuhua Li; Molly D Wong; Bin Zheng; Xizeng Wu; Hong Liu
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

3.  Low dose high energy x-ray in-line phase sensitive imaging prototype: Investigation of optimal geometric conditions and design parameters.

Authors:  Muhammad U Ghani; Aimin Yan; Molly D Wong; Yuhua Li; Liqiang Ren; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2015       Impact factor: 1.535

4.  Dose and detectability improvements with high energy phase sensitive x-ray imaging in comparison to low energy conventional imaging.

Authors:  Molly Donovan Wong; Aimin Yan; Muhammad Ghani; Yuhua Li; Laurie Fajardo; Xizeng Wu; Hong Liu
Journal:  Phys Med Biol       Date:  2014-04-15       Impact factor: 3.609

5.  Image quality and dose efficiency of high energy phase sensitive x-ray imaging: phantom studies.

Authors:  Molly Donovan Wong; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2014       Impact factor: 1.535

6.  Quantitative investigation of the edge enhancement in in-line phase contrast projections and tomosynthesis provided by distributing microbubbles on the interface between two tissues: a phantom study.

Authors:  Di Wu; Molly Donovan Wong; Yuhua Li; Laurie Fajardo; Bin Zheng; Xizeng Wu; Hong Liu
Journal:  Phys Med Biol       Date:  2017-11-21       Impact factor: 3.609

7.  Adaptation of a clustered lumpy background model for task-based image quality assessment in x-ray phase-contrast mammography.

Authors:  Adam M Zysk; Jovan G Brankov; Miles N Wernick; Mark A Anastasio
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

8.  Soft tissue small avascular tumor imaging with x-ray phase-contrast micro-CT in-line holography.

Authors:  Yakov Nesterets; Tim Gureyev; Andrew Stevenson; Andrew Pogany; Steve Wilkins; Russell Kincaid; Hongwei Ye; Levon Vogelsang; Edward Lipson; Ioana Coman; Sylvain Fourmaux; Jean-Claude Kieffer; Andrzej Krol
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2008

9.  The effects of x-ray beam hardening on detective quantum efficiency and radiation dose.

Authors:  Molly Donovan Wong; Xizeng Wu; Hong Liu
Journal:  J Xray Sci Technol       Date:  2011       Impact factor: 1.535

10.  The Potential of Utilizing Mid-Energy X-Rays for In-Line Phase Sensitive Breast Cancer Imaging.

Authors:  F H Omoumi; M U Ghani; M D Wong; Y Li; B Zheng; A Yan; P A Jenkins; X Wu; H Liu
Journal:  Biomed Spectrosc Imaging       Date:  2020-12-28
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.