Literature DB >> 14638542

Long-term follow-up of incomplete stent apposition in patients who received sirolimus-eluting stent for de novo coronary lesions: an intravascular ultrasound analysis.

Muzaffer Degertekin1, Patrick W Serruys, Kengo Tanabe, Chi Hang Lee, J Edouardo Sousa, Antonio Colombo, Marie-Claude Morice, Jurgen M R Ligthart, Pim J de Feyter.   

Abstract

BACKGROUND: Incomplete stent apposition (ISA) has been previously documented after sirolimus-eluting stent (SES) implantation. The aim of this study was to investigate the long-term intravascular ultrasound (IVUS) findings of ISA in patients who received SES. METHODS AND
RESULTS: A total of 13 patients who received SES and showed ISA at follow-up IVUS (follow-up I) were investigated. IVUS was performed on all of these patients 12 months later (follow-up II). Quantitative ISA area measurement was also performed at follow-up I and II. No vascular remodeling was observed in the vessel segment with ISA; external elastic membrane area was 19.4+/-6.6 versus 19.5+/-6.4 mm2 at follow-up I and II, respectively. There was also no significant change in external elastic membrane area between vessel segment with ISA and without ISA (+1.5% versus -3.0%, respectively; P=0.27) at late follow-up. The ISA area, either including (2.5+/-1.7 versus 3.8+/-6.3 mm2; P=NS) or excluding (2.5+/-1.8 versus 2.4+/-1.7 mm2; P=NS) a single patient with aneurysm formation, was not significantly different between follow-up I and II. One patient manifested a coronary aneurysm in the stented segment at late follow-up that was probably present at the initial follow-up but masked by thrombus. It was successfully treated with a covered stent. All patients were asymptomatic, and no patient experienced late thrombotic occlusion.
CONCLUSIONS: Vessel dimensions and area of ISA did not change over time, except for 1 coronary aneurysm that became apparent. ISA after implantation of a SES was not associated with adverse events at late follow-up.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14638542     DOI: 10.1161/01.CIR.0000103666.25660.77

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  13 in total

1.  Comparison of vascular remodeling in patients treated with sirolimus-versus zotarolimus-eluting stent following acute myocardial infarction.

Authors:  Ki-Woon Kang; Young-Guk Ko; Dong-Ho Shin; Jung-Sun Kim; Byeong-Keuk Kim; Donghoon Choi; Yangsoo Jang; Myeong-Ki Hong
Journal:  Clin Cardiol       Date:  2011-12-12       Impact factor: 2.882

Review 2.  Applications of grayscale and radiofrequency intravascular ultrasound to image atherosclerotic plaque.

Authors:  Somjot S Brar; Gary S Mintz; Akiko Maehara; Gregg W Stone
Journal:  J Nucl Cardiol       Date:  2010-10       Impact factor: 5.952

3.  Drug deposition in coronary arteries with overlapping drug-eluting stents.

Authors:  Farhad Rikhtegar; Elazer R Edelman; Ufuk Olgac; Dimos Poulikakos; Vartan Kurtcuoglu
Journal:  J Control Release       Date:  2016-07-16       Impact factor: 9.776

4.  Comparison of intravascular ultrasonic imaging with versus without incomplete stent apposition at follow-up after drug-eluting stent implantation.

Authors:  Juying Qian; Feng Zhang; Hongyi Wu; Bing Fan; Lei Ge; Junbo Ge
Journal:  Int J Cardiovasc Imaging       Date:  2007-08-17       Impact factor: 2.357

5.  Parent vessel size and curvature strongly influence risk of incomplete stent apposition in enterprise intracranial aneurysm stent coiling.

Authors:  R S Heller; A M Malek
Journal:  AJNR Am J Neuroradiol       Date:  2011-08-04       Impact factor: 3.825

6.  Extensive late-acquired incomplete stent apposition after sirolimus-eluting stent implantation.

Authors:  Hyun-Chul Choi; Seung-Ho Hur; Chang-Wook Nam; Yun-Kyeong Cho; Hyungseop Kim; Nam-Hee Park; Yoon-Nyun Kim; Kwon-Bae Kim
Journal:  Korean Circ J       Date:  2010-01-27       Impact factor: 3.243

7.  Assessment of the relation between IVUS measurements and clinical outcome in elderly patients after sirolimus-eluting stent implantation for de novo coronary lesions.

Authors:  Run Du; Rui Yan Zhang; Qi Zhang; Yu Hang Shi; Jian Hu; Zhen Kun Yang; Feng Hua Ding; Jian Sheng Zhang; Wei Feng Shen
Journal:  Int J Cardiovasc Imaging       Date:  2012-01-06       Impact factor: 2.357

8.  Late stent malapposition with marked positive vascular remodeling observed only at the site of drug-eluting stents after multivessel coronary stenting.

Authors:  Yusuke Watanabe; Ryuta Asano; Noriko Hata; Kanki Inoue; Itaru Takamisawa; Atsushi Seki; Masaru Aikawa; Tetsuya Tobaru; Kazuhiko Misu; Nobuo Iguchi; Masatoshi Nagayama; Hiroyuki Watanabe; Morimasa Takayama; Jun Umemura; Tetsuya Sumiyoshi
Journal:  Heart Vessels       Date:  2009-07-22       Impact factor: 2.037

9.  Vascular ultrasound for atherosclerosis imaging.

Authors:  Chris L de Korte; Hendrik H G Hansen; Anton F W van der Steen
Journal:  Interface Focus       Date:  2011-06-01       Impact factor: 3.906

10.  Angiographic spontaneous pseudo-resolution of a coronary artery aneurysm after implantation of a sirolimus-eluting stent.

Authors:  Yong-Woo Choi; Hong-Seok Lim; Seung-Jea Tahk; Myeong-Ho Yoon; Hyoung-Mo Yang; Xiong-Jie Jin
Journal:  Korean J Intern Med       Date:  2016-01-28       Impact factor: 2.884

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.