[Measurement of scotopic pupils: comparison of scotic pupil measurements using a colvard pupilometer and the slitlamp green light test].

W Sekundo1, G Nietgen, J C Schmidt.   

Abstract

PURPOSE: The aim of the study was to compare the accuracy and the reproducibility of scotopic pupil measurements using two different methods. PATIENTS AND METHODS: We developed a simple test to measure scotopic pupil diameters using a narrow slit of light through a green filter of the Haag-Streit slitlamp. A total of 100 eyes from 50 refractive surgery candidates were prospectively examined by 2 independent investigators using both the Colvard pupilometer (Oasis/USA) and Sekundo's slitlamp green light test. Results were compared using all pairwise multiple comparison procedures (Student-Newman-Keuls method, SigmaStat/Jandel Scientific). The mean age of the patients was 36.3 years with a male:female ratio of 16:34. The colour of the iris was considered blue for 36 individuals and brown for the remainder.
RESULTS: Reproducibility: the right eye mean pupil diameter using Sekundo's method was 6.4 mm (+/-0.9) measured by both the first (Slit 1R) and the second (Slit 2R) investigator. The left eye mean pupil diameter was 6.5 mm (+/-1) (Slit 1L) and 6.35 mm (+/-1) (Slit 2L), respectively. The following measurements were obtained with the Colvard pupillometer: (Colv1R)=6.25+/-0.85 mm, (Colv2R)=5.99+/-1 mm, (Colv1L)=6.15+/-0.91 mm, (Colv2L)=6.05+/-0.9 mm. There were no statistically significant differences between the two investigators within the same method of examination. Comparability:The data of both investigators were combined to form 4 groups: (Slit R)=6.4+/-0.96 mm, (Slit L)=6.4+/-1 mm, (Colv R)=6.05+/-0.9 mm, (Colv L)=6.1+/-0.93 mm. There was a significant difference ( p=0.024) between the two methods for both right and left eyes. Particularly for large pupils, the slitlamp green light measurements were significantly higher (up to 0.35 mm) than those with Colvard's device.
CONCLUSIONS: The slitlamp green light test provides similar reproducibility for the measurement of the scotopic pupil diameter as the commercially available Colvard pupilometer.

Entities:  

Mesh:

Year:  2003        PMID: 14618354     DOI: 10.1007/s00347-003-0806-2

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  7 in total

1.  Scotopic measurement of normal pupils: Colvard versus Video Vision Analyzer infrared pupillometer.

Authors:  E M Schnitzler; M Baumeister; T Kohnen
Journal:  J Cataract Refract Surg       Date:  2000-06       Impact factor: 3.351

2.  Comparison of the pupil card and pupillometer in measuring pupil size.

Authors:  Mihai Pop; Yves Payette; Emma Santoriello
Journal:  J Cataract Refract Surg       Date:  2002-02       Impact factor: 3.351

3.  Pupillary size and responsiveness. Implications for selection of a bifocal intraocular lens.

Authors:  D D Koch; S W Samuelson; E A Haft; L M Merin
Journal:  Ophthalmology       Date:  1991-07       Impact factor: 12.079

4.  The relationship of visual acuity, refractive error, and pupil size after radial keratotomy.

Authors:  J T Holladay; M J Lynn; G O Waring; M Gemmill; G C Keehn; B Fielding
Journal:  Arch Ophthalmol       Date:  1991-01

5.  Preoperative measurement of scotopic pupil dilation using an office pupillometer.

Authors:  M Colvard
Journal:  J Cataract Refract Surg       Date:  1998-12       Impact factor: 3.351

6.  Use of a digital infrared pupillometer to assess patient suitability for refractive surgery.

Authors:  Emanuel S Rosen; Christa L Gore; Daniel Taylor; Deepak Chitkara; Frank Howes; Elizabeth Kowalewski
Journal:  J Cataract Refract Surg       Date:  2002-08       Impact factor: 3.351

7.  Night vision after excimer laser photorefractive keratectomy: haze and halos.

Authors:  D P O'Brart; C P Lohmann; F W Fitzke; S E Smith; M G Kerr-Muir; J Marshall
Journal:  Eur J Ophthalmol       Date:  1994 Jan-Mar       Impact factor: 1.922

  7 in total
  3 in total

1.  [Classification of biomedical research reports as a reference for evidence-based medicine in ophthalmology. A survey considering as example the journal Der Ophthalmologe].

Authors:  H P N Scholl; M Fleckenstein; T U Krohne; F G Holz
Journal:  Ophthalmologe       Date:  2005-12       Impact factor: 1.059

2.  [Influence of incision parameters on astigmatism during implantation of phakic-6-mm-iris-claw intraocular lenses].

Authors:  W Sekundo; M Schneider; A Tietjen
Journal:  Ophthalmologe       Date:  2004-03       Impact factor: 1.059

3.  [A prospective intraindividual comparison between laser in situ keratomileusis and laser subepithelial keratectomy for myopia. 1-year follow-up results].

Authors:  A Tietjen; C Müller; W Sekundo
Journal:  Ophthalmologe       Date:  2008-10       Impact factor: 1.059

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.