Vascular calcification in ex vivo carotid specimens: precision and accuracy of measurements with multi-detector row CT.

Udo Hoffmann1, Dylan C Kwait, Jason Handwerker, Raymond Chan, Glenn Lamuraglia, Thomas J Brady.   

Abstract

PURPOSE: To test the accuracy and precision of multi-detector row computed tomography (CT)-derived measurements of vascular calcification in ex vivo human carotid endarterectomy (CEA) specimens.
MATERIALS AND METHODS: Sixteen ex vivo CEA specimens were imaged with multi-detector row CT. Multi-detector row CT-derived calcium scoring algorithms (ie, mineral mass and volume score) were compared with the mass and volume of ashed remnants of the CEA specimens. Bland-Altman analysis was performed to assess the mean (ie, bias) and SD (ie, precision) of differences between multi-detector row CT- and ashing-derived measurements. In addition, conventional Agatston score, volume score, mineral mass, and modified Agatston score were calculated for each specimen by using a number of scanning protocols. Images were obtained at a section thickness of 1.25 mm by using different tube energy settings and tube currents. Specimens were also imaged at different section thicknesses with fixed combinations of tube energy and tube current. To compare measurement variability of scoring methods, coefficients of variation for all protocols were calculated.
RESULTS: Both mean multi-detector row CT-derived mineral mass and mean ashing-derived mineral mass were 0.129 g +/- 0.173 (r = 0.99, P <.001). Mean multi-detector row CT- and ashing-derived volumes were 339.94 mm3 +/- 395.4 and 39.48 mm3 +/- 55.76, respectively (r = 0.95, P <.001). Measurement bias relative to the reference ashing values was high (2,800.0%) for volume score and low (2.58%) for mineral mass. Measurement precision was 0.6% for volume score and greater than 0.0005% for mineral mass. Mean coefficients of variation for all CT protocols were 5.0% +/- 4.2 and 4.9% +/- 4.2 for mineral mass and modified Agatston score, respectively, and 16.0% +/- 9.2 and 14.5% +/- 3.9 for conventional Agatston and volume scores, respectively (P <.001).
CONCLUSION: Compared with the conventional volume score, multi-detector row CT-derived mineral mass is a less biased and more precise measurement of the mineral content of nonmoving ex vivo CEA specimens. Mineral mass and modified Agatston score are more reproducible than conventional volume and Agatston scores. Copyright RSNA, 2003

Entities:  

Mesh:

Year:  2003        PMID: 14526093     DOI: 10.1148/radiol.2292021016

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  15 in total

Review 1.  Understanding the genetics of coronary artery disease through the lens of noninvasive imaging.

Authors:  Eunice Yang; Jose D Vargas; David A Bluemke
Journal:  Expert Rev Cardiovasc Ther       Date:  2012-01

2.  Mineral volume and morphology in carotid plaque specimens using high-resolution MRI and CT.

Authors:  Ronald L Wolf; Suzanne L Wehrli; Andra M Popescu; John H Woo; Hee Kwon Song; Alexander C Wright; Emile R Mohler; John D Harding; Eric L Zager; Ronald M Fairman; Michael A Golden; Omaida C Velazquez; Jeffrey P Carpenter; Felix W Wehrli
Journal:  Arterioscler Thromb Vasc Biol       Date:  2005-06-09       Impact factor: 8.311

3.  A novel model to test accuracy and reproducibility of MDCT scan protocols for coronary calcium in vivo.

Authors:  Michael Rosol; Karina Sachdev; Christian N Enzweiler; Dylan C Kwait; Ryan Millea; James Titus; Jason Handwerker; Stephan Wicky; Stephen Achenbach; Thomas J Brady; Udo Hoffmann
Journal:  Int J Cardiovasc Imaging       Date:  2005-12-20       Impact factor: 2.357

4.  Correlation between carotid bifurcation calcium burden on non-enhanced CT and percentage stenosis, as confirmed by digital subtraction angiography.

Authors:  B Sarikaya; B Lohman; A M McKinney; S Gadani; M Irfan; L Lucato
Journal:  Br J Radiol       Date:  2011-09-06       Impact factor: 3.039

5.  Dual energy computed tomography quantification of carotid plaques calcification: comparison between monochromatic and polychromatic energies with pathology correlation.

Authors:  Lorenzo Mannelli; Lawrence MacDonald; Marcello Mancini; Marina Ferguson; William P Shuman; Monica Ragucci; Serena Monti; Dongxiang Xu; Chun Yuan; Lee M Mitsumori
Journal:  Eur Radiol       Date:  2014-12-24       Impact factor: 5.315

6.  Kidney stone volume estimation from computerized tomography images using a model based method of correcting for the point spread function.

Authors:  Xinhui Duan; Jia Wang; Mingliang Qu; Shuai Leng; Yu Liu; Amy Krambeck; Cynthia McCollough
Journal:  J Urol       Date:  2012-07-21       Impact factor: 7.450

7.  Effect of different hydroxyapatite incorporation methods on the structural and biological properties of porous collagen scaffolds for bone repair.

Authors:  Alan J Ryan; John P Gleeson; Amos Matsiko; Emmet M Thompson; Fergal J O'Brien
Journal:  J Anat       Date:  2014-11-20       Impact factor: 2.610

8.  Changes in measured size of atherosclerotic plaque calcifications in dual-energy CT of ex vivo carotid endarterectomy specimens: effect of monochromatic keV image reconstructions.

Authors:  Lorenzo Mannelli; Lee M Mitsumori; Marina Ferguson; Dongxiang Xu; Baocheng Chu; Kelley R Branch; William P Shuman; Chun Yuan
Journal:  Eur Radiol       Date:  2012-08-21       Impact factor: 5.315

9.  Association between calcification of the cervical carotid artery bifurcation and white matter ischemia.

Authors:  N F Fanning; T D Walters; A J Fox; S P Symons
Journal:  AJNR Am J Neuroradiol       Date:  2006-02       Impact factor: 3.825

10.  Calcium scoring using 64-slice MDCT, dual source CT and EBT: a comparative phantom study.

Authors:  Jaap M Groen; Marcel J W Greuter; R Vliegenthart; C Suess; B Schmidt; F Zijlstra; M Oudkerk
Journal:  Int J Cardiovasc Imaging       Date:  2007-11-23       Impact factor: 2.357

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.