A novel model to test accuracy and reproducibility of MDCT scan protocols for coronary calcium in vivo.

Michael Rosol1, Karina Sachdev, Christian N Enzweiler, Dylan C Kwait, Ryan Millea, James Titus, Jason Handwerker, Stephan Wicky, Stephen Achenbach, Thomas J Brady, Udo Hoffmann.   

Abstract

OBJECTIVES: We compared the accuracy and reliability of prospectively triggered, retrospectively ECG gated, and non-gated CT image reconstruction for measurements of coronary artery calcification (CAC) in vivo using a novel animal model.
MATERIALS AND METHODS: In six Yorkshire farm pigs, prefabricated chains of cortical bone fragments were sutured over the epicardial bed of the major coronary arteries. Using a 4-slice MDCT scanner, each animal was imaged with two different protocols: sequential acquisition with prospective ECG triggering, and spiral acquisition with retrospectively ECG gated image reconstruction- non-gated reconstructions were also generated from these latter scans. Two independent observers measured the 'Agatston score' (AS), the calcified volume (CV), and mineral mass (MM). To calculate accuracy of MM measurements the ash weight of the burned bone fragments was compared to MDCT derived MM.
RESULTS: Six pigs successfully underwent surgery and CT imaging (mean heart rate: 86+/-12 bpm). MM measurements from prospectively ECG triggered CT sequential scans were more accurate (p<0.02) and reproducible (p=0.05) than sequential CT scans without ECG triggering or spiral acquisition using retrospective ECG gating.
CONCLUSIONS: At high heart rates prospective ECG triggered image reconstruction is more accurate and reproducible for CAC scoring than retrospective ECG gated reconstruction and non-gated reconstruction.

Entities:  

Mesh:

Year:  2005        PMID: 16374527     DOI: 10.1007/s10554-005-6535-7

Source DB:  PubMed          Journal:  Int J Cardiovasc Imaging        ISSN: 1569-5794            Impact factor:   2.357


  27 in total

1.  Noninvasive coronary angiography by retrospectively ECG-gated multislice spiral CT.

Authors:  S Achenbach; S Ulzheimer; U Baum; M Kachelriess; D Ropers; T Giesler; W Bautz; W G Daniel; W A Kalender; W Moshage
Journal:  Circulation       Date:  2000-12-05       Impact factor: 29.690

2.  Coronary artery calcium measurement: agreement of multirow detector and electron beam CT.

Authors:  C R Becker; T Kleffel; A Crispin; A Knez; J Young; U J Schoepf; R Haberl; M F Reiser
Journal:  AJR Am J Roentgenol       Date:  2001-05       Impact factor: 3.959

Review 3.  Radiation dose in computed tomography of the heart.

Authors:  Richard L Morin; Thomas C Gerber; Cynthia H McCollough
Journal:  Circulation       Date:  2003-02-18       Impact factor: 29.690

4.  Coronary artery motion during the cardiac cycle and optimal ECG triggering for coronary artery imaging.

Authors:  B Lu; S S Mao; N Zhuang; H Bakhsheshi; H Yamamoto; J Takasu; S C Liu; M J Budoff
Journal:  Invest Radiol       Date:  2001-05       Impact factor: 6.016

5.  Quantification of coronary artery calcium using ultrafast computed tomography.

Authors:  A S Agatston; W R Janowitz; F J Hildner; N R Zusmer; M Viamonte; R Detrano
Journal:  J Am Coll Cardiol       Date:  1990-03-15       Impact factor: 24.094

6.  Probabilistic model for prediction of angiographically defined obstructive coronary artery disease using electron beam computed tomography calcium score strata.

Authors:  L F Bielak; J A Rumberger; P F Sheedy; R S Schwartz; P A Peyser
Journal:  Circulation       Date:  2000-07-25       Impact factor: 29.690

7.  Chemical and physicochemical studies on the mineral deposits of the human atherosclerotic aorta.

Authors:  K Schmid; W O McSharry; C H Pameijer; J P Binette
Journal:  Atherosclerosis       Date:  1980-10       Impact factor: 5.162

8.  Sensitivity to detect small coronary artery calcium lesions with varying slice thickness using electron beam tomography.

Authors:  Songshou Mao; Janis Child; Sivi Carson; Steve C K Liu; Ronald J Oudiz; Matthew J Budoff
Journal:  Invest Radiol       Date:  2003-03       Impact factor: 6.016

9.  Ultra-low-dose coronary artery calcium screening using multislice CT with retrospective ECG gating.

Authors:  Tobias F Jakobs; Bernd J Wintersperger; Peter Herzog; Thomas Flohr; Christoph Suess; Andreas Knez; Maximilian F Reiser; Christoph R Becker
Journal:  Eur Radiol       Date:  2003-05-21       Impact factor: 5.315

10.  Electron-beam CT: use of a calibration phantom to reduce variability in calcium quantitation.

Authors:  C H McCollough; R B Kaufmann; B M Cameron; D J Katz; P F Sheedy; P A Peyser
Journal:  Radiology       Date:  1995-07       Impact factor: 11.105

View more
  1 in total

1.  Coronary calcium mass scores measured by identical 64-slice MDCT scanners are comparable: a cardiac phantom study.

Authors:  Hildebrand Dijkstra; Marcel J W Greuter; Jaap M Groen; Rozemarijn Vliegenthart-Proença; Klaasjan W K Renema; Frank de Lange; Matthijs Oudkerk
Journal:  Int J Cardiovasc Imaging       Date:  2009-09-19       Impact factor: 2.357

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.