Clinical safety of magnetic resonance imaging early after coronary artery stent placement.

Thomas C Gerber1, Panayotis Fasseas, Ryan J Lennon, Venkata U Valeti, Christopher P Wood, Jerome F Breen, Peter B Berger.   

Abstract

OBJECTIVES: Our aim was to examine the rate of adverse cardiac events in patients undergoing magnetic resonance imaging (MRI) <8 weeks after coronary stent placement.
BACKGROUND: The risk of coronary stent thrombosis from dislodgement due to MRI early after stent placement is not well defined. Manufacturers recommend postponing MRI studies until eight weeks after coronary stent placement.
METHODS: We analyzed the Mayo Clinic Rochester Percutaneous Coronary Intervention Database and examined records of 111 patients who underwent MRI <8 weeks after coronary stent placement treated with aspirin and a thienopyridine. Occurrence of death, myocardial infarction (MI), and repeat revascularization within 30 days of MRI were recorded.
RESULTS: Magnetic resonance imaging (1.5 tesla) was performed within a median of 18 days (range, 0 to 54 days) after coronary stent placement. Four noncardiac deaths occurred, and three patients had repeat revascularization procedures. Stent thrombosis did not occur (95% confidence interval, 0% to 3.3%).
CONCLUSIONS: Magnetic resonance imaging <8 weeks after coronary stent placement appears to be safe, and the risk of cardiac death or MI due to stent thrombosis is low. Postponing MRI does not appear to be necessary.

Entities:  

Mesh:

Year:  2003        PMID: 14522498     DOI: 10.1016/s0735-1097(03)00993-8

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  13 in total

1.  ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents.

Authors:  W Gregory Hundley; David A Bluemke; J Paul Finn; Scott D Flamm; Mark A Fogel; Matthias G Friedrich; Vincent B Ho; Michael Jerosch-Herold; Christopher M Kramer; Warren J Manning; Manesh Patel; Gerald M Pohost; Arthur E Stillman; Richard D White; Pamela K Woodard
Journal:  Circulation       Date:  2010-05-17       Impact factor: 29.690

Review 2.  ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents.

Authors:  W Gregory Hundley; David A Bluemke; J Paul Finn; Scott D Flamm; Mark A Fogel; Matthias G Friedrich; Vincent B Ho; Michael Jerosch-Herold; Christopher M Kramer; Warren J Manning; Manesh Patel; Gerald M Pohost; Arthur E Stillman; Richard D White; Pamela K Woodard
Journal:  J Am Coll Cardiol       Date:  2010-06-08       Impact factor: 24.094

3.  Safety of cardiovascular magnetic resonance in patients with cardiovascular implants and devices.

Authors:  S K Prasad; D J Pennell
Journal:  Heart       Date:  2004-11       Impact factor: 5.994

Review 4.  Cardiovascular magnetic resonance: structure, function, perfusion, and viability.

Authors:  David C Isbell; Christopher M Kramer
Journal:  J Nucl Cardiol       Date:  2005 May-Jun       Impact factor: 5.952

5.  Contrast-enhanced cardiovascular magnetic resonance in the hyperacute phase of ST-elevation myocardial infarction.

Authors:  Eric Larose; Julie Côté; Josep Rodés-Cabau; Bernard Noël; Gerald Barbeau; Edith Bordeleau; Santiago Miró; Bernard Brochu; Robert Delarochellière; Olivier F Bertrand
Journal:  Int J Cardiovasc Imaging       Date:  2009-03-15       Impact factor: 2.357

6.  MRI demonstration of acute myocardial infarction due to posttraumatic coronary artery dissection.

Authors:  Savas M Tepe; James F Glockner; Paul Julsrud
Journal:  Int J Cardiovasc Imaging       Date:  2005-07-23       Impact factor: 2.357

7.  Cardiac magnetic resonance imaging safety following percutaneous coronary intervention.

Authors:  Jason W Curtis; Donna C Lesniak; James H Wible; Pamela K Woodard
Journal:  Int J Cardiovasc Imaging       Date:  2013-04-27       Impact factor: 2.357

Review 8.  Biocompatible Polymer Materials with Antimicrobial Properties for Preparation of Stents.

Authors:  Kateřina Škrlová; Kateřina Malachová; Alexandra Muñoz-Bonilla; Dagmar Měřinská; Zuzana Rybková; Marta Fernández-García; Daniela Plachá
Journal:  Nanomaterials (Basel)       Date:  2019-10-31       Impact factor: 5.076

9.  Coronary computed tomography and magnetic resonance imaging.

Authors:  Birgit Kantor; Eike Nagel; Paul Schoenhagen; Jörg Barkhausen; Thomas C Gerber
Journal:  Curr Probl Cardiol       Date:  2009-04       Impact factor: 5.200

10.  3-Tesla MRI for the evaluation of myocardial viability: a comparative study with 1.5-Tesla MRI.

Authors:  G Ligabue; F Fiocchi; S Ferraresi; A Barbieri; R Rossi; M G Modena; R Romagnoli; P Torricelli
Journal:  Radiol Med       Date:  2008-07-09       Impact factor: 3.469

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.