Reducing the radiation dose during excretory urography: flat-panel silicon x-ray detector versus computed radiography.

M Zähringer1, V Hesselmann, O Schulte, K F Kamm, W Braun, G Haupt, B Krug, K Lackner.   

Abstract

OBJECTIVE: The purpose of the study was to examine the possibilities for reducing radiation exposure in uroradiology using digital flat-panel silicon X-ray detector radiography. We compared the subjectively determined image quality of abdominal radiographs and urograms obtained on a digital flat-panel detector radiography system with those obtained on a computed radiography system. SUBJECTS AND METHODS. Fifty patients who had a clinical indication for urography underwent unenhanced abdominal imaging that was alternately performed using flat-panel silicon X-ray detector radiography or computed radiography. For patients who required a second radiograph with contrast medium, the examination modality was changed to avoid exposing the patients to excess radiation. The images obtained on flat-panel X-ray detector radiography were obtained at half the radiation dose of the images obtained on computed radiography (800 speed vs 400 speed). The resulting 50 pairs of images were interpreted by four independent observers who rated the detectability of structures of bone and the efferent urinary tract relevant to diagnosis and compared the image quality.
RESULTS: At half the radiation dose, digital flat-panel X-ray detector radiography provided equivalent image quality of the liver and spleen, lumbar vertebrae 2 and 5, pelvis, and psoas margin on abdominal radiographs. The image quality obtained with digital flat-panel X-ray detector radiography of the kidneys, the hollow cavities of the upper efferent urinary tract, and the urinary bladder was judged to be statistically better than those obtained with computed radiography.
CONCLUSION: With half the exposure dose of computed radiography, the flat-panel X-ray detector produced urograms with an image quality equivalent to or better than computed radiography.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14500204     DOI: 10.2214/ajr.181.4.1810931

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  2 in total

1.  Reducing dose in urography while maintaining image quality-a comparison of storage phosphor plates and a flat-panel detector.

Authors:  Margareta Jansson; Håkan Geijer; Jan Persliden; Torbjörn Andersson
Journal:  Eur Radiol       Date:  2005-04-27       Impact factor: 5.315

2.  Modeling the performance characteristics of computed radiography (CR) systems.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  IEEE Trans Med Imaging       Date:  2010-03       Impact factor: 10.048

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.