Comparison of single-photon and dual-energy x-ray absorptiometry of the radius.

D Nelson1, M Feingold, E Mascha, M Kleerekoper.   

Abstract

We compared dual-energy x-ray absorptiometry (DXA) for measurement of the radius with the conventional single-photon absorptiometry (SPA) method. To evaluate reproducibility, 34 healthy male and female subjects were measured twice each by both methods. While the instruments measured bone mineral content (BMC) similarly, projected area was consistently lower by SPA and therefore bone mineral density (BMD) was higher by an average of 10%. We report similar coefficients of variation for the two methods, which are 0.8% (SPA) and 0.7% (DXA) for BMC (P = 0.71) and 0.8% (SPA) and 1.4% (DXA) for BMD (P = 0.02). We also evaluated the relationship between the methods with data from 196 clinic patients who were measured once each on both instruments. The BMD measurements from the two instruments were highly correlated (r = 0.97) in these patients, which permits the conversion of databases from SPA to DXA-equivalents. We conclude that DXA can replace SPA for radial bone densitometry.

Entities:  

Mesh:

Year:  1992        PMID: 1422300     DOI: 10.1016/0169-6009(92)90801-j

Source DB:  PubMed          Journal:  Bone Miner        ISSN: 0169-6009


  4 in total

Review 1.  Current methods and advances in bone densitometry.

Authors:  G Guglielmi; C C Gluer; S Majumdar; B A Blunt; H K Genant
Journal:  Eur Radiol       Date:  1995       Impact factor: 5.315

2.  Radial bone density and breast cancer risk in white and African-American women.

Authors:  D A Nelson; L L Darga; M S Simon; R K Severson
Journal:  Osteoporos Int       Date:  2004-07       Impact factor: 4.507

3.  Accuracy and precision study in vitro for peripheral quantitative computed tomography.

Authors:  M Takada; K Engelke; S Hagiwara; S Grampp; H K Genant
Journal:  Osteoporos Int       Date:  1996       Impact factor: 4.507

4.  Densitometry of the radius using single and dual energy absorptiometry.

Authors:  K G Faulkner; M R McClung; M S Schmeer; L A Roberts; K W Gaither
Journal:  Calcif Tissue Int       Date:  1994-03       Impact factor: 4.333

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.