Comparison of Sentinel and Bactec blood culture systems.

M Stevens1, H Patel, A Walters, K Burch, A Jay, N Dowling, C J Mitchell, R A Swann, A T Willis, D C Shanson.   

Abstract

AIMS: To evaluate the Sentinel automated blood culture system and to compare its performance with that of Bactec.
METHODS: The Sentinel blood culture system was evaluated in three centres. The performance of the system was assessed in comparison with the routine blood culture method used in these centres, the Bactec system.
RESULTS: Blood culture sets (n = 2180) consisting of Sentinel aerobic and anaerobic, and Bactec aerobic and anaerobic bottles yielded 218 (10%) clinically important isolates. One hundred and fifty five (71%) of the isolates were detected by both systems; 35 (16%) were detected by Sentinel only; and 28 (13%) by Bactec only. For the duration of the evaluation, the Sentinel system was deliberately configured so that it was impossible to detect positive results during the first 12 hours. The times to positivity after the first 12 hours were similar. Data gathered during and subsequent to the evaluation have been used by the manufacturer to refine the algorithm so that positive results can be detected at a minimum of 2.25 hours.
CONCLUSIONS: After a period of familiarization the Sentinel system was considered easy to use. Sentinel is a useful addition to the methods available for the detection of bacteria in blood cultures.

Entities:  

Mesh:

Year:  1992        PMID: 1401215      PMCID: PMC495112          DOI: 10.1136/jcp.45.9.815

Source DB:  PubMed          Journal:  J Clin Pathol        ISSN: 0021-9746            Impact factor:   3.411


  4 in total

Review 1.  Blood culture technique: current controversies.

Authors:  D C Shanson
Journal:  J Antimicrob Chemother       Date:  1990-04       Impact factor: 5.790

2.  Clinical comparison of a new automated infrared blood culture system with the BACTEC 460 system.

Authors:  D Jungkind; J Millan; S Allen; J Dyke; E Hill
Journal:  J Clin Microbiol       Date:  1986-02       Impact factor: 5.948

3.  Microbiological comparison of a new infra-red blood culture system (Bactec NR 600) and a radiometric system (Bactec 460).

Authors:  J E Corkill; K Rimmer
Journal:  Med Lab Sci       Date:  1987-04

4.  Controlled evaluation of the volume of blood cultured in detection of bacteremia and fungemia.

Authors:  J H Tenney; L B Reller; S Mirrett; W L Wang; M P Weinstein
Journal:  J Clin Microbiol       Date:  1982-04       Impact factor: 5.948

  4 in total
  3 in total

1.  Sentinel and Bactec blood culture systems.

Authors:  R P Cooke; C T Jenkins
Journal:  J Clin Pathol       Date:  1993-03       Impact factor: 3.411

2.  Detection of bacteraemia by the continuously monitoring BacT/Alert system.

Authors:  G T Kennedy; J G Barr; C Goldsmith
Journal:  J Clin Pathol       Date:  1995-10       Impact factor: 3.411

3.  Development of o.a.s.i.s., a new automated blood culture system in which detection is based on measurement of bottle headspace pressure changes.

Authors:  C M Stevens; D Swaine; C Butler; A H Carr; A Weightman; C R Catchpole; D E Healing; T S Elliott
Journal:  J Clin Microbiol       Date:  1994-07       Impact factor: 5.948

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.