Intravascular ultrasound imaging of the coronary arteries: an in vitro evaluation of measurement of area of the lumen and atheroma characterisation.

M H Anderson1, I A Simpson, D Katritsis, M J Davies, D E Ward.   

Abstract

OBJECTIVE: To assess the accuracy of measurement of area of the lumen, and sensitivity, and specificity of detection of atheroma in coronary arteries in vitro with a commercially available 20 MHz intravascular ultrasound system.
SETTING: A teaching hospital department of cardiology with the support of the department of cardiovascular pathology. PROCEDURE: 10 segments of coronary artery were removed from cadaver hearts. Intravascular ultrasound imaging was performed at fixed levels and the vessels were then sectioned and photographed before histological preparation. An independent blinded observer measured luminal area and assessed the presence of atheroma on the intravascular ultrasound images of 76 vessel sections (304 quadrants). The sensitivity and specificity of detection of atheroma was assessed in comparison with the histologically prepared sections. Luminal areas from intravascular ultrasound, photographs of cross sections of the vessels and histological sections were compared with the technique of limits of agreement.
RESULTS: Overall 36% of the 304 quadrants studied histologically had identifiable atheroma. Intravascular ultrasound sensitivity for atheroma was 0.593 and the specificity was 0.839. The positive predictive value was 0.674, and the relative risk 3.139. Values for area of the vessel lumen were on average 9.4 mm2 (confidence interval (CI) 8.6-10.2 mm2) larger than those measured from photographs and 10.7 (CI 9.8-11.6 mm2) larger than those measured from the histological sections.
CONCLUSIONS: The intravascular ultrasound system assessed in this study significantly overestimated coronary vessel luminal area and had low sensitivity and specificity for detection of atheroma. Improvements in image resolution are required before this system can provide useful information on coronary artery size and morphology.

Entities:  

Mesh:

Year:  1992        PMID: 1389758      PMCID: PMC1025070          DOI: 10.1136/hrt.68.9.276

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  20 in total

1.  Clinical percutaneous imaging of coronary anatomy using an over-the-wire ultrasound catheter system.

Authors:  J M Hodgson; S P Graham; A D Savakus; S G Dame; D N Stephens; P S Dhillon; D Brands; H Sheehan; M J Eberle
Journal:  Int J Card Imaging       Date:  1989

2.  Optimized ultrasound imaging catheters for use in the vascular system.

Authors:  R J Crowley; P L von Behren; L A Couvillon; D E Mai; J E Abele
Journal:  Int J Card Imaging       Date:  1989

3.  Intravascular ultrasound cross-sectional arterial imaging before and after balloon angioplasty in vitro.

Authors:  J M Tobis; J A Mallery; J Gessert; J Griffith; D Mahon; M Bessen; M Moriuchi; L McLeay; M McRae; W L Henry
Journal:  Circulation       Date:  1989-10       Impact factor: 29.690

4.  Arterial wall characteristics determined by intravascular ultrasound imaging: an in vitro study.

Authors:  E J Gussenhoven; C E Essed; C T Lancée; F Mastik; P Frietman; F C van Egmond; J Reiber; H Bosch; H van Urk; J Roelandt
Journal:  J Am Coll Cardiol       Date:  1989-10       Impact factor: 24.094

5.  Intravascular ultrasound estimation of arterial stenosis.

Authors:  N G Pandian; A Kreis; T O'Donnell
Journal:  J Am Soc Echocardiogr       Date:  1989 Nov-Dec       Impact factor: 5.251

6.  Discrepancies between cineangiographic and postmortem findings in patients with coronary artery disease and recent myocardial revascularization.

Authors:  C M Grondin; I Dyrda; A Pasternac; L Campeau; M G Bourassa; J Lespérance
Journal:  Circulation       Date:  1974-04       Impact factor: 29.690

7.  Variability in the analysis of coronary arteriograms.

Authors:  T A DeRouen; J A Murray; W Owen
Journal:  Circulation       Date:  1977-02       Impact factor: 29.690

8.  Real-time intravascular ultrasound imaging in humans.

Authors:  N G Pandian; A Kreis; A Weintraub; A Motarjeme; M Desnoyers; J M Isner; M Konstam; D N Salem; V Millen
Journal:  Am J Cardiol       Date:  1990-06-01       Impact factor: 2.778

9.  Coronary artery imaging with intravascular high-frequency ultrasound.

Authors:  B N Potkin; A L Bartorelli; J M Gessert; R F Neville; Y Almagor; W C Roberts; M B Leon
Journal:  Circulation       Date:  1990-05       Impact factor: 29.690

Review 10.  Restenosis after coronary angioplasty. Potential biologic determinants and role of intimal hyperplasia.

Authors:  M W Liu; G S Roubin; S B King
Journal:  Circulation       Date:  1989-06       Impact factor: 29.690

View more
  5 in total

1.  Importance of calibration for diameter and area determination by intravascular ultrasound.

Authors:  P Stähr; H J Rupprecht; T Voigtländer; P Kearney; R Erbel; L Koch; S Krass; R Brennecke; J Meyer
Journal:  Int J Card Imaging       Date:  1996-12

2.  Factors influencing the intravascular ultrasound dimensions and areas measured using mechanical transducers.

Authors:  V Bhargava; W F Penny; A Arbab-Zadeh
Journal:  Int J Card Imaging       Date:  1996-12

Review 3.  Intracoronary ultrasound: current state of the art.

Authors:  P P Kearney; I R Starkey; G R Sutherland
Journal:  Br Heart J       Date:  1995-05

4.  Atherosclerotic coronary lesions with inadequate compensatory enlargement have smaller plaque and vessel volumes: observations with three dimensional intravascular ultrasound in vivo.

Authors:  C von Birgelen; G S Mintz; E A de Vrey; T Kimura; J J Popma; S G Airiian; M B Leon; M Nobuyoshi; P W Serruys; P J de Feyter
Journal:  Heart       Date:  1998-02       Impact factor: 5.994

5.  Adjunctive intra-coronary imaging for the assessment of coronary artery disease.

Authors:  Nikunj Shah; Bassey Ussen; Michael Mahmoudi
Journal:  JRSM Cardiovasc Dis       Date:  2016-07-08
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.