Literature DB >> 1376899

Clinical evaluation of the safety of repetitive intraoperative defibrillation threshold testing.

R Frame1, R Brodman, S Furman, S G Kim, J Roth, K Ferrick, I Hollinger, J Gross, J D Fisher.   

Abstract

One goal of the initial implantation procedure for a cardioverter defibrillator is determination of the configuration and patch location with the lowest defibrillation threshold (DFT). To determine the safety of multiple defibrillation tests, an analysis of the intraoperative defibrillation threshold tests (DFTT) in our patients was performed. In 84 patients, the mean number of DFT trials was 5.27; the mean number of joules received was 275.0. The maximum number of shocks in one implant procedure was 50 for a total of 4,895 joules without complications. Four patients received 30 or more DFT shocks without complication. There were two complications related directly to the DFTT: one patient with severe noninflammatory cardiomyopathy developed electromechanical dissociation and was subsequently resuscitated and survived; the second patient with severe triple vessel coronary artery disease suffered an intraoperative myocardial infarction during testing and eventually died 22 days postoperatively. All patients received an ICD unit; six patients had DFTs of greater than 20 joules. Based on our experience, we followed the clinical status (heart rate, blood pressure, ECG changes, fluid status, total anesthesia time) during the DFTT to determine the extent and duration of our testing protocol. Multiple shocks due to repositioning of the leads in a stable patient should not prohibit extensive testing as adverse consequences do not appear to be cumulative.

Entities:  

Mesh:

Year:  1992        PMID: 1376899     DOI: 10.1111/j.1540-8159.1992.tb03077.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  8 in total

1.  Effect of defibrillation threshold testing-induced ventricular fibrillation on renal function.

Authors:  John H Shin; Chotikorn Khunnawat; Jose Baez-Escudero; Bradley P Knight; John F Beshai
Journal:  J Interv Card Electrophysiol       Date:  2013-10-11       Impact factor: 1.900

2.  Correlation of acute and chronic defibrillation threshold with upper limit of vulnerability determined in normal sinus rhythm.

Authors:  U Birgersdotter-Green; K Undesser; O Fujimura; G K Feld; R M Kass; W J Mandel; C T Peter; P S Chen
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

3.  Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices.

Authors:  Saman Nazarian; Rozann Hansford; Amir A Rahsepar; Valeria Weltin; Diana McVeigh; Esra Gucuk Ipek; Alan Kwan; Ronald D Berger; Hugh Calkins; Albert C Lardo; Michael A Kraut; Ihab R Kamel; Stefan L Zimmerman; Henry R Halperin
Journal:  N Engl J Med       Date:  2017-12-28       Impact factor: 91.245

4.  A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices.

Authors:  Saman Nazarian; Rozann Hansford; Ariel Roguin; Dorith Goldsher; Menekhem M Zviman; Albert C Lardo; Brian S Caffo; Kevin D Frick; Michael A Kraut; Ihab R Kamel; Hugh Calkins; Ronald D Berger; David A Bluemke; Henry R Halperin
Journal:  Ann Intern Med       Date:  2011-10-04       Impact factor: 25.391

5.  Clinical utility and safety of a protocol for noncardiac and cardiac magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5 tesla.

Authors:  Saman Nazarian; Ariel Roguin; Menekhem M Zviman; Albert C Lardo; Timm L Dickfeld; Hugh Calkins; Robert G Weiss; Ronald D Berger; David A Bluemke; Henry R Halperin
Journal:  Circulation       Date:  2006-09-11       Impact factor: 29.690

6.  Defibrillation Testing During Defibrillator Implantation.

Authors:  Dietmar Bänsch
Journal:  Arrhythm Electrophysiol Rev       Date:  2012-09

7.  2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement on optimal implantable cardioverter-defibrillator programming and testing.

Authors:  Bruce L Wilkoff; Laurent Fauchier; Martin K Stiles; Carlos A Morillo; Sana M Al-Khatib; Jesœs Almendral; Luis Aguinaga; Ronald D Berger; Alejandro Cuesta; James P Daubert; Sergio Dubner; Kenneth A Ellenbogen; N A Mark Estes; Guilherme Fenelon; Fermin C Garcia; Maurizio Gasparini; David E Haines; Jeff S Healey; Jodie L Hurtwitz; Roberto Keegan; Christof Kolb; Karl-Heinz Kuck; Germanas Marinskis; Martino Martinelli; Mark McGuire; Luis G Molina; Ken Okumura; Alessandro Proclemer; Andrea M Russo; Jagmeet P Singh; Charles D Swerdlow; Wee Siong Teo; William Uribe; Sami Viskin; Chun-Chieh Wang; Shu Zhang
Journal:  J Arrhythm       Date:  2016-02-01

8.  Defibrillation testing of the implantable cardioverter defibrillator: when, how, and by whom?

Authors:  Luis A Pires
Journal:  Indian Pacing Electrophysiol J       Date:  2007-08-01
  8 in total