Color and conventional image-directed Doppler ultrasonography: accuracy and sources of error in quantitative blood flow measurements.

C Ranke1, P Hendrickx, U Roth, F Brassel, A Creutzig, K Alexander.   

Abstract

Accuracy of two systems--conventional (DRF 400, Diasonics) and color-coded (Angiodynograph, Quantum/Phillips) image-directed Doppler ultrasonography--was investigated using an in vitro model that generated both monophasic and triphasic pulsatile flow patterns. Estimated and actual blood volume flow rates showed good correlations, but the sampling with a hand-held transducer led to wide variations in measurement error for the conventional (-69.2% to 50%) and the color-coded (-79.3% to 265.7%) systems. By performing multiple measurements, one could improve accuracy considering only the maximal values of a series instead of the mean values. Accuracy was impaired by interposed muscular or fatty tissue due to false low time-average velocity measurements caused by a loss of Doppler signal. Comparison of both systems revealed significant differences between pulsatility index values (p less than 0.001), blood flow velocities (p less than 0.001), and blood volume flow rates (p less than 0.05 for program flow, p less than 0.001 for manual and automatic flow program of the color-coded system).

Mesh:

Year:  1992        PMID: 1313832     DOI: 10.1002/jcu.1870200305

Source DB:  PubMed          Journal:  J Clin Ultrasound        ISSN: 0091-2751            Impact factor:   0.910


  9 in total

Review 1.  [Angiology update].

Authors:  C Ranke; H J Trappe
Journal:  Med Klin (Munich)       Date:  1999-05-15

2.  Fetal dynamic phase-contrast MR angiography using ultrasound gating and comparison with Doppler ultrasound measurements.

Authors:  B P Schoennagel; J Yamamura; F Kording; R Fischer; P Bannas; G Adam; H Kooijman; C Ruprecht; K Fehrs; M Tavares de Sousa
Journal:  Eur Radiol       Date:  2019-01-07       Impact factor: 5.315

3.  The clinical implications of adopting new criteria for the grading of internal carotid artery stenosis.

Authors:  Nicole Curtis; Martin Necas; Matthew Versteeg
Journal:  Australas J Ultrasound Med       Date:  2018-02-14

4.  Magnetic Resonance Venous Volume Measurements in Peripheral Artery Disease (from ELIMIT).

Authors:  Hassan Kamran; Vijay Nambi; Smita Negi; Eric Y Yang; Changyi Chen; Salim S Virani; Panos Kougias; Alan B Lumsden; Joel D Morrisett; Christie M Ballantyne; Gerd Brunner
Journal:  Am J Cardiol       Date:  2016-08-13       Impact factor: 2.778

5.  Foetal blood flow measured using phase contrast cardiovascular magnetic resonance--preliminary data comparing 1.5 T with 3.0 T.

Authors:  Beverly Tsai-Goodman; Meng Yuan Zhu; Mashael Al-Rujaib; Mike Seed; Christopher K Macgowan
Journal:  J Cardiovasc Magn Reson       Date:  2015-04-18       Impact factor: 5.364

6.  Prediction of restenosis based on hemodynamical markers in revascularized femoro-popliteal arteries during leg flexion.

Authors:  Can Gökgöl; Nicolas Diehm; Lorenz Räber; Philippe Büchler
Journal:  Biomech Model Mechanobiol       Date:  2019-06-13

7.  Feasibility of quantification of the distribution of blood flow in the normal human fetal circulation using CMR: a cross-sectional study.

Authors:  Mike Seed; Joshua F P van Amerom; Shi-Joon Yoo; Bahiyah Al Nafisi; Lars Grosse-Wortmann; Edgar Jaeggi; Michael S Jansz; Christopher K Macgowan
Journal:  J Cardiovasc Magn Reson       Date:  2012-11-26       Impact factor: 5.364

8.  Fetal circulation in left-sided congenital heart disease measured by cardiovascular magnetic resonance: a case-control study.

Authors:  Bahiyah Al Nafisi; Joshua F P van Amerom; Jonathan Forsey; Edgar Jaeggi; Lars Grosse-Wortmann; Shi-Joon Yoo; Christopher K Macgowan; Mike Seed
Journal:  J Cardiovasc Magn Reson       Date:  2013-07-27       Impact factor: 5.364

Review 9.  Brain perfusion imaging in neonates.

Authors:  Jérôme Baranger; Olivier Villemain; Matthias Wagner; Mariella Vargas-Gutierrez; Mike Seed; Olivier Baud; Birgit Ertl-Wagner; Julien Aguet
Journal:  Neuroimage Clin       Date:  2021-07-14       Impact factor: 4.881

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.