Literature DB >> 12918186

[Reliability of home monitoring with event-recording compared with polysomnography in infants].

Sabine de Nardi1, Ekkehart Paditz, Thomas Erler, Andreas Gruntzke.   

Abstract

AIM OF THE STUDY: Although reliable recognition of hypoxemia, apnea, bradycardia and tachycardia is absolutely necessary for home monitoring, many commercially available home monitors have not been sufficiently tested for their sensitivity. The purpose of this study was to determine the reliability of the home monitor VitaGuard 3000 by comparing it with the manual evaluation of full polysomnography (polysomnographic system Alice 3). PATIENT AND METHODS: 20 infants (11 males; 9 females) aged between 5 and 40 weeks (12.2 +/- 3.5 weeks, median 10.5) and with a gestational age between 29 and 41 weeks (37.7 +/- 6.1 weeks, median 39.0) were tested using both full polysomnography and, simultaneously, the home monitor VitaGuard 3000. The results were evaluated manually and compared.
RESULTS: The monitor system detected 7/51 central apneas, 6/260 desaturations and 7/18 tachycardias. The sensitivity was 13.72% for apnea, 4.23% for desaturation and 38.80% for tachycardia. The reliability of the home monitor for detecting apnea, desaturation and tachycardia was therefore insufficient.
CONCLUSION: A polysomnographic reliability test should be mandatory for all home monitoring systems prior to commercial introduction.

Entities:  

Mesh:

Year:  2003        PMID: 12918186     DOI: 10.1007/bf03040435

Source DB:  PubMed          Journal:  Wien Klin Wochenschr        ISSN: 0043-5325            Impact factor:   1.704


  13 in total

1.  False alarm rates of three third-generation pulse oximeters in PACU, ICU and IABP patients.

Authors:  Norbert O Lutter; Sabine Urankar; Steffi Kroeber
Journal:  Anesth Analg       Date:  2002-01       Impact factor: 5.108

2.  Telemetry as a new concept in long term monitoring of SIDS-risk infant.

Authors:  A F Aly; D Afchine; P Esser; M Joos; H J Niewerth; A Wiater; M Meier; D Padeken; A Pericas; D Schwartmann; T Weber; V Wendrix; M Wirtz
Journal:  Eur J Med Res       Date:  2000-01-26       Impact factor: 2.175

3.  "Motion-resistant" pulse oximetry: a comparison of new and old models.

Authors:  Steven J Barker
Journal:  Anesth Analg       Date:  2002-10       Impact factor: 5.108

4.  [The Graz SIDS Risk Questionnaire: prospective study of 6,000 infants].

Authors:  C Einspieler; W N Löscher; R Kurz; K Rosanelli; H Rosegger; I Bachler; E M Klug; F Reiterer; R Schenkeli; R Kerbl
Journal:  Klin Padiatr       Date:  1992 Mar-Apr       Impact factor: 1.349

5.  Home monitoring for infants at high risk for the sudden infant death syndrome.

Authors:  Y Sivan; A Kornecki; A Baharav; N Glaser; Z Spirer
Journal:  Isr J Med Sci       Date:  1997-01

Review 6.  [Home monitoring of infants at risk of sudden infant death: suggestions for reconsideration of current practice].

Authors:  C F Poets
Journal:  Wien Klin Wochenschr       Date:  2000-03-10       Impact factor: 1.704

7.  [Evaluation of the cardiorespiratory monitor SpiroGuard C for infants. Improved registration of respiratory events by new sensors and intelligent alarm management system].

Authors:  M S Urschitz; F Eichler; C Popow; O S Ipsiroglu
Journal:  Wien Klin Wochenschr       Date:  2000-03-10       Impact factor: 1.704

8.  Pneumogram recordings and cerebral computed tomography as predictors of the severity of infant apnea.

Authors:  J S Deacon; M T Hynan; C Blaser
Journal:  J Perinatol       Date:  1993 Nov-Dec       Impact factor: 2.521

9.  Reliability of conventional and new pulse oximetry in neonatal patients.

Authors:  William W Hay; Donna J Rodden; Shannon M Collins; Diane L Melara; Kathy A Hale; Lucy M Fashaw
Journal:  J Perinatol       Date:  2002 Jul-Aug       Impact factor: 2.521

10.  Detection of hyperoxaemia in neonates: data from three new pulse oximeters.

Authors:  B Bohnhorst; C S Peter; C F Poets
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-11       Impact factor: 5.747

View more
  1 in total

1.  [Reliability of home monitoring with event recording compared with polysomnography in infants].

Authors:  Herwig von Nettelhorst; Robert Downes
Journal:  Wien Klin Wochenschr       Date:  2004-04-30       Impact factor: 1.704

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.