A revised mathematical treatment of the calibration line intercept has been published for in vivo bone lead measurements using 109Cd-based K-shell x-ray fluorescence. The revised calibration line treatment prompts changes, presented herein, to the method for calculating the measurement uncertainty.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12884934 DOI: 10.1088/0031-9155/48/13/314