Literature DB >> 12853517

Usefulness of new automated cutting needle for tissue-core biopsy of lung nodules under CT fluoroscopic guidance.

Takuji Yamagami1, Shigeharu Iida, Takeharu Kato, Osamu Tanaka, Shogo Toda, Daishiro Kato, Tsunehiko Nishimura.   

Abstract

OBJECTIVES: The goal of our study was to evaluate the efficacy and safety of a new type of automated cutting needle for tissue-core lung biopsy under real-time CT fluoroscopic guidance.
DESIGN: Observational
MATERIALS AND METHODS: One hundred ten percutaneous needle tissue-core lung biopsies were performed with the Auto Surecut needle (Create Medic; Yokohama, Japan) under CT fluoroscopic guidance, and the specimens obtained underwent histopathologic evaluation. A final diagnosis was confirmed by independent surgical pathology.
RESULTS: The rate of success for the diagnosis for specimens that were adequate for histopathologic analysis was 94.5% (104 of 110 specimens). The sensitivity, specificity, and accuracy in diagnosing malignancy were 95.1%, 100%, and 96.2%, respectively. A specific cell type could be characterized in 95.2% of those 104 lesions (99 lesions; malignant, 76 lesions; benign, 23 lesions). The specific cell type was precisely diagnosed, and was confirmed after surgery in 65 malignant lesions and 23 benign lesions. The biopsy-induced complications encountered were pneumothorax in 34.5% (38 of 110 patients) and hemoptysis in 6.4% (7 of 110 patients). No patient had a serious complication.
CONCLUSION: CT fluoroscopy-guided lung biopsy using the Auto Surecut needle provides a high degree of diagnostic accuracy, allows for the specific characterization of lung nodules, and can be performed safely.

Entities:  

Mesh:

Year:  2003        PMID: 12853517     DOI: 10.1378/chest.124.1.147

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  16 in total

1.  Accuracy and complications in computed tomography fluoroscopy-guided needle biopsies of lung masses.

Authors:  Siri Lagethon Heck; Peter Blom; Audun Berstad
Journal:  Eur Radiol       Date:  2006-03-16       Impact factor: 5.315

2.  Comparison of fully automated and semi-automated biopsy needles for lung biopsy under CT fluoroscopic guidance.

Authors:  R Yoshimatsu; T Yamagami; O Tanaka; H Miura; T Tanaka; T Suzuki; T Nishimura
Journal:  Br J Radiol       Date:  2011-08-09       Impact factor: 3.039

3.  Effectiveness of combined use of imprint cytological and histological examination in CT-guided tissue-core biopsy.

Authors:  Takuji Yamagami; Rika Yoshimatsu; Kenji Kajiwara; Masaki Ishikawa; Tomohiro Matsumoto; Hideaki Kakizawa; Naoyuki Toyoda; Terumitsu Hasebe; Kazuo Awai
Journal:  Eur Radiol       Date:  2014-02-14       Impact factor: 5.315

4.  Diagnostic feasibility and safety of CT-guided core biopsy for lung nodules less than or equal to 8 mm: A single-institution experience.

Authors:  Ying-Yueh Chang; Chun-Ku Chen; Yi-Chen Yeh; Mei-Han Wu
Journal:  Eur Radiol       Date:  2017-09-07       Impact factor: 5.315

5.  CT fluoroscopy-guided lung biopsy versus conventional CT-guided lung biopsy: a prospective controlled study to assess radiation doses and diagnostic performance.

Authors:  Ga Ram Kim; Jin Hur; Sang Min Lee; Hye-Jeong Lee; Yoo Jin Hong; Ji Eun Nam; Hua Sun Kim; Young Jin Kim; Byoung Wook Choi; Tae Hoon Kim; Kyu Ok Choe
Journal:  Eur Radiol       Date:  2010-08-22       Impact factor: 5.315

6.  Diagnostic performance of percutaneous lung biopsy using automated biopsy needles under CT-fluoroscopic guidance for ground-glass opacity lesions.

Authors:  T Yamagami; R Yoshimatsu; H Miura; K Yamada; A Takahata; T Matsumoto; T Hasebe
Journal:  Br J Radiol       Date:  2013-02       Impact factor: 3.039

7.  Role of CT-guided percutaneous lung biopsy in diagnosis of pulmonary fungal infection in patients with hematologic diseases.

Authors:  Ji-Min Shi; Zhen Cai; He Huang; Xiu-Jin Ye; Jing-Song He; Wan-Zhuo Xie; Jie Zhang; Xian-Yong Zhou; Yi Luo; Yu Lin; Li Li; Wei-Yan Zheng; Guo-Qing Wei; Mao-Fang Lin
Journal:  Int J Hematol       Date:  2009-05-27       Impact factor: 2.490

8.  Evaluation of the efficacy and safety of percutaneous biopsy of lung.

Authors:  Sean E McSweeney; Kevin N O'Regan; Patrick D Mc Laughlin; Lee Crush; Michael M Maher
Journal:  Open Respir Med J       Date:  2012-09-06

9.  Combined fluoroscopy- and CT-guided transthoracic needle biopsy using a C-arm cone-beam CT system: comparison with fluoroscopy-guided biopsy.

Authors:  Joo Yeon Cheung; Yookyung Kim; Sung Shine Shim; Soo Mee Lim
Journal:  Korean J Radiol       Date:  2011-01-03       Impact factor: 3.500

10.  CT-Guided Transthoracic Biopsy of Lung Lesions Using a Non-Coaxial Biopsy Needle Technique: CT Characteristics Predictive for Diagnostic Accuracy and Pneumothorax.

Authors:  Barbara Geeroms; Lesley Cockmartin; Johan Coolen; Adriana Dubbeldam; Johny Verschakelen; Ani Nikoghosyan; Walter De Wever
Journal:  J Belg Soc Radiol       Date:  2021-06-30       Impact factor: 1.894

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.