Literature DB >> 12837593

On-line strong cation exchange micro-HPLC-ESI-MS/MS for protein identification and process optimization.

Thierry Le Bihan1, Henry S Duewel, Daniel Figeys.   

Abstract

We have developed an on-line strong cation exchange (SCX)-ESI-MS/MS platform for the rapid identification of proteins contained in mixtures. This platform consists of a SCX precolumn followed by a nanoflow SCX column on-line with an electrospray ion trap mass spectrometer. We also used this platform to study the dynamics of peptide separation/extraction by SCX, in particular to understand the parameters affecting the performance of SCX in multidimensional chromatography. For example, we have demonstrated that the buffer typically used for tryptic digestion of protein mixtures can have a detrimental effect on the chromatographic behaviour of peptides during SCX separations, thereby affecting certain peptide quantitation approaches that rely on reproducible peptide fractionation. We have also demonstrated that band broadening results when a step (discontinuous) gradient approach is used to displace peptides from the SCX precolumn, reducing the separation power of SCX in multidimensional chromatography. In contrast, excellent chromatographic peak shapes are observed when a defined (continuous) gradient is used. Finally, using a tryptic digest of a protein extract derived from human K562 cells, we observed that larger molecular weight peptides are identified using this on-line SCX approach compared to the more conventional reverse phase (RP) LC/MS approach. Both methods used in tandem complement each other and can lead to a greater number of peptide identifications from a given sample.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12837593     DOI: 10.1016/s1044-0305(03)00208-3

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  17 in total

Review 1.  Membrane proteins and proteomics: un amour impossible?

Authors:  V Santoni; M Molloy; T Rabilloud
Journal:  Electrophoresis       Date:  2000-04       Impact factor: 3.535

Review 2.  The dynamic range of protein expression: a challenge for proteomic research.

Authors:  G L Corthals; V C Wasinger; D F Hochstrasser; J C Sanchez
Journal:  Electrophoresis       Date:  2000-04       Impact factor: 3.535

Review 3.  Analysis of proteins and proteomes by mass spectrometry.

Authors:  M Mann; R C Hendrickson; A Pandey
Journal:  Annu Rev Biochem       Date:  2001       Impact factor: 23.643

Review 4.  The application of mass spectrometry to membrane proteomics.

Authors:  Christine C Wu; John R Yates
Journal:  Nat Biotechnol       Date:  2003-03       Impact factor: 54.908

5.  Selective detection of membrane proteins without antibodies: a mass spectrometric version of the Western blot.

Authors:  David Arnott; Adrianne Kishiyama; Elizabeth A Luis; Sarah G Ludlum; James C Marsters; John T Stults
Journal:  Mol Cell Proteomics       Date:  2002-02       Impact factor: 5.911

6.  Automation of nanoscale microcapillary liquid chromatography-tandem mass spectrometry with a vented column.

Authors:  Lawrence J Licklider; Carson C Thoreen; Junmin Peng; Steven P Gygi
Journal:  Anal Chem       Date:  2002-07-01       Impact factor: 6.986

7.  Evaluation of multidimensional chromatography coupled with tandem mass spectrometry (LC/LC-MS/MS) for large-scale protein analysis: the yeast proteome.

Authors:  Junmin Peng; Joshua E Elias; Carson C Thoreen; Larry J Licklider; Steven P Gygi
Journal:  J Proteome Res       Date:  2003 Jan-Feb       Impact factor: 4.466

8.  Comprehensive two-dimensional high-performance liquid chromatography for the isolation of overexpressed proteins and proteome mapping.

Authors:  G J Opiteck; S M Ramirez; J W Jorgenson; M A Moseley
Journal:  Anal Biochem       Date:  1998-05-01       Impact factor: 3.365

9.  The focusing positions of polypeptides in immobilized pH gradients can be predicted from their amino acid sequences.

Authors:  B Bjellqvist; G J Hughes; C Pasquali; N Paquet; F Ravier; J C Sanchez; S Frutiger; D Hochstrasser
Journal:  Electrophoresis       Date:  1993-10       Impact factor: 3.535

10.  Direct analysis of protein complexes using mass spectrometry.

Authors:  A J Link; J Eng; D M Schieltz; E Carmack; G J Mize; D R Morris; B M Garvik; J R Yates
Journal:  Nat Biotechnol       Date:  1999-07       Impact factor: 54.908

View more
  4 in total

1.  Gradient chromatofocusing-mass spectrometry: a new technique in protein analysis.

Authors:  Lian Shan; James A Hribar; Xiang Zhou; David J Anderson
Journal:  J Am Soc Mass Spectrom       Date:  2008-05-15       Impact factor: 3.109

2.  An interaction network of the mammalian COP9 signalosome identifies Dda1 as a core subunit of multiple Cul4-based E3 ligases.

Authors:  Michael Hans Olma; Marcia Roy; Thierry Le Bihan; Izabela Sumara; Sarah Maerki; Brett Larsen; Manfredo Quadroni; Matthias Peter; Mike Tyers; Lionel Pintard
Journal:  J Cell Sci       Date:  2009-04-01       Impact factor: 5.285

3.  High Sensitivity Quantitative Proteomics Using Automated Multidimensional Nano-flow Chromatography and Accumulated Ion Monitoring on Quadrupole-Orbitrap-Linear Ion Trap Mass Spectrometer.

Authors:  Paolo Cifani; Alex Kentsis
Journal:  Mol Cell Proteomics       Date:  2017-08-18       Impact factor: 5.911

Review 4.  Advanced proteomic liquid chromatography.

Authors:  Fang Xie; Richard D Smith; Yufeng Shen
Journal:  J Chromatogr A       Date:  2012-07-09       Impact factor: 4.759

  4 in total