Flow injection of the lock mass standard for accurate mass measurement in electrospray ionization time-of-flight mass spectrometry coupled with liquid chromatography.

Laurence Charles1.   

Abstract

A method of flow injection of the lock mass for accurate mass measurement using electrospray ionization time-of-flight mass spectrometry is described. The reference compound is introduced in the chromatographic effluent via a six-port valve placed post-column, prior to the split connector. Flow injection is performed in such a way that the reference elution peak is superimposed in the total ion current and partially overlaps that of the investigated analyte, allowing independent ionization of the two compounds and thus accurate mass measurement with no ion suppression effects. Different lock mass molecules can be injected in a single analytical run to target various analytes. The performance of this methodology is demonstrated in both isocratic and gradient liquid chromatography modes. The molecular ion of the flow-injected lock mass could also be used as a reference for mass measurement of the in-source fragments of the analytes. Good mass accuracy, within 4 mDa of the theoretical values, was obtained. Copyright 2003 John Wiley & Sons, Ltd.

Year:  2003        PMID: 12820201     DOI: 10.1002/rcm.1060

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  5 in total

1.  Utilizing Microchip Capillary Electrophoresis Electrospray Ionization for Hydrogen Exchange Mass Spectrometry.

Authors:  William A Black; Bradley B Stocks; J Scott Mellors; John R Engen; J Michael Ramsey
Journal:  Anal Chem       Date:  2015-06-02       Impact factor: 6.986

2.  Assigning product ions from complex MS/MS spectra: the importance of mass uncertainty and resolving power.

Authors:  Lekha Sleno; Dietrich A Volmer; Alan G Marshall
Journal:  J Am Soc Mass Spectrom       Date:  2005-02       Impact factor: 3.109

3.  Sub-part-per-million precursor and product mass accuracy for high-throughput proteomics on an electron transfer dissociation-enabled orbitrap mass spectrometer.

Authors:  Craig D Wenger; Graeme C McAlister; Qiangwei Xia; Joshua J Coon
Journal:  Mol Cell Proteomics       Date:  2010-02-02       Impact factor: 5.911

4.  Correcting systematic bias and instrument measurement drift with mzRefinery.

Authors:  Bryson C Gibbons; Matthew C Chambers; Matthew E Monroe; David L Tabb; Samuel H Payne
Journal:  Bioinformatics       Date:  2015-08-04       Impact factor: 6.937

5.  High-resolution extracted ion chromatography, a new tool for metabolomics and lipidomics using a second-generation orbitrap mass spectrometer.

Authors:  Albert Koulman; Gary Woffendin; Vinod K Narayana; Helen Welchman; Catharina Crone; Dietrich A Volmer
Journal:  Rapid Commun Mass Spectrom       Date:  2009-05       Impact factor: 2.419

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.