Interference caused by the contents of serum separator tubes in the Vitros CRP assay.

Chia-Yu Chang1, Jin-Ying Lu, Tzu-I Chien, Jau-Tsuen Kao, Mei-Chun Lin, Pei-Chun Shih, Shiao-Ni Yan.   

Abstract

BACKGROUND: We have observed discrepancies between C-reactive protein (CRP) measured in serum prepared from a serum separator tube (SST) and that obtained from a plain tube, when using the Vitros CRP assay. Our study aimed at elucidating the cause of these discrepancies.
METHODS: Eighty-seven specimens from hospitalized patients with various types of inflammatory disease were analysed using a fixed-point immuno-rate method on a Vitros CRP slide. The serum was prepared simultaneously in both vacuum and SSTs. We also performed mixing tests by adding 47 samples of serum prepared from plain tubes to SSTs and incubating for 15 min before CRP analysis.
RESULTS: Lower values of CRP were found in serum prepared from plain tubes than in serum from SSTs. Addition of serum prepared from plain tubes to SSTs and incubating for 15 min increased the CRP values significantly. The ratio of CRP measured in serum prepared from plain tubes and from SSTs did not differ significantly from the ratio obtained when serum was prepared in a plain tube then added to an SST. DISCUSSION: We propose that SSTs can adsorb some macromolecules that form complexes with CRP. The addition of SST gel to serum results in the release of CRP molecules from these complexes, which enhances the antigen-antibody reaction on the Vitros CRP slide and increases the measured CRP concentrations.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12803837     DOI: 10.1258/000456303321610556

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  4 in total

1.  Evaluation of BD Vacutainer SST™ II Plus tubes for common tumor marker tests by Roche Diagnostics Modular E 170 analyzer.

Authors:  Zhiyan Li; Zhenru Feng; Cunling Yan; Rong Yan
Journal:  J Clin Lab Anal       Date:  2010       Impact factor: 2.352

2.  Evaluation of BD Vacutainer SST™ II plus tubes for special proteins testing.

Authors:  Zhiyan Li; Cunling Yan; Rong Yan; Xinzhi Zheng; Zhenru Feng
Journal:  J Clin Lab Anal       Date:  2011       Impact factor: 2.352

3.  Interferences in immunoassay.

Authors:  Jill Tate; Greg Ward
Journal:  Clin Biochem Rev       Date:  2004-05

4.  Abnormal gel flotation in a patient with apperant pneumonia diagnosis: a case report.

Authors:  Fethullah Gerin; Dilber Coban Ramazan; Ozgur Baykan; Onder Sirikci; Goncagul Haklar
Journal:  Biochem Med (Zagreb)       Date:  2014-02-15       Impact factor: 2.313

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.