Literature DB >> 12794601

The development and use of a chest phantom for optimizing scanning techniques on a variety of low-dose helical computed tomography devices.

Yoshihisa Muramatsu1, Yukihiro Tsuda, Yoshimasa Nakamura, Mitsuru Kubo, Toshiyuki Takayama, Kouzou Hanai.   

Abstract

PURPOSE: To develop a chest phantom for determining the optimal scan conditions for chest computed tomography (CT) screening.
METHODS: The basic structure of the phantom is an arms-elevated-positioned anthropomorphic chest phantom. The internal structure includes simulated tumors (which are assumed to be the target lesions of chest CT screening examinations) placed at the levels of the lung apex, the bifurcation of the trachea, and the lung base of both simulated lungs. The image contrast of the simulated tumors is taken as the difference in CT value and is specified as 1 of 2 target values: Delta270 and Delta100. An opening for placement of a dosimeter is provided on the central axis of the phantom. Initial field tests were conducted focusing on the scan conditions for chest CT screening.
RESULTS: Images similar to those obtained in chest CT screening examinations and clearly showing pathologic changes were obtained. The dose measurement at the center was 2.0 mGy. The diameters of the simulated tumors that could be detected were 6 mm for Delta270 and 10 mm for Delta100.
CONCLUSION: The use of this phantom makes it possible to determine the optimal scan conditions for chest CT screening based on objective evaluation criteria.

Entities:  

Mesh:

Year:  2003        PMID: 12794601     DOI: 10.1097/00004728-200305000-00012

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  12 in total

1.  Data analysis of low dose multislice helical CT scan in orbital trauma.

Authors:  Ji-Wei Wang; Chong Tang; Bo-Rong Pan
Journal:  Int J Ophthalmol       Date:  2012-06-18       Impact factor: 1.779

2.  Low-dose CT screening for lung cancer with automatic exposure control: phantom study.

Authors:  Shiho Gomi; Yoshihisa Muramatsu; Shinsuke Tsukagoshi; Masahiro Suzuki; Ryutaro Kakinuma; Ryosuke Tsuchiya; Noriyuki Moriyama
Journal:  Radiol Phys Technol       Date:  2008-07-11

3.  A computer-simulated liver phantom (virtual liver phantom) for multidetector computed tomography evaluation.

Authors:  Yoshinori Funama; Kazuo Awai; Osamu Miyazaki; Yoshiharu Nakayama; Da Liu; Taiga Goto; Yasuyuki Yamashita; Shinichi Hori
Journal:  Eur Radiol       Date:  2005-10-20       Impact factor: 5.315

4.  Automated volumetry of pulmonary nodules on multidetector CT: influence of slice thickness, reconstruction algorithm and tube current. Preliminary results.

Authors:  A R Larici; M L Storto; M Torge; M Mereu; F Molinari; F Maggi; L Bonomo
Journal:  Radiol Med       Date:  2008-02-25       Impact factor: 3.469

5.  Identification and characterization of focal ground-glass opacity in the lungs by high-resolution CT using thin-section multidetector helical CT: experimental study using a chest CT phantom.

Authors:  Duo Liu; Kazuo Awai; Yoshinori Funama; Seitaro Oda; Takeshi Nakaura; Yumi Yanaga; Masahiro Hatemura; Koichi Kawanaka; Yasuyuki Yamashita
Journal:  Radiat Med       Date:  2008-01-31

6.  Characterization of a novel anthropomorphic plastinated lung phantom.

Authors:  Sungwon Yoon; Robert W Henry; Donna M Bouley; N Robert Bennett; Rebecca Fahrig
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

7.  Detection of pure ground-glass nodules in the lung by low-dose multi-detector computed tomography, with use of an iterative reconstruction method: a comparison with conventional image reconstruction by the filtered back-projection method.

Authors:  Shiho Akashita; Yasuhiko Tachibana; Kentaro Sakamaki; Keiji Sogawa; Tomio Inoue
Journal:  Jpn J Radiol       Date:  2015-01-01       Impact factor: 2.374

8.  A method for evaluating the performance of computer-aided detection of pulmonary nodules in lung cancer CT screening: detection limit for nodule size and density.

Authors:  Hajime Kobayashi; Masaki Ohkubo; Akihiro Narita; Janaka C Marasinghe; Kohei Murao; Toru Matsumoto; Shusuke Sone; Shinichi Wada
Journal:  Br J Radiol       Date:  2017-01-03       Impact factor: 3.039

9.  Detection of ground-glass opacities by use of hybrid iterative reconstruction (iDose) and low-dose 256-section computed tomography: a phantom study.

Authors:  Ko Higuchi; Michinobu Nagao; Yoshio Matsuo; Shunya Sunami; Takeshi Kamitani; Mikako Jinnouchi; Masato Yonezawa; Yuzo Yamasaki; Hidetake Yabuuchi; Masamitsu Hatkenaka; Hiroshi Honda
Journal:  Radiol Phys Technol       Date:  2013-02-12

10.  Observer-independent nodule-detectability index for low-dose lung cancer screening CT: a pilot study.

Authors:  Masaki Ohkubo; Shinichi Wada; Satoshi Kanai; Kazuhiro Ishikawa; Janaka C Marasinghe; Toru Matsumoto
Journal:  Radiol Phys Technol       Date:  2013-06-09
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.