Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging.

Dirk Verellen1, Guy Soete, Nadine Linthout, Swana Van Acker, Patsy De Roover, Vincent Vinh-Hung, Jan Van de Steene, Guy Storme.   

Abstract

BACKGROUND AND
PURPOSE: The aim of this study is to investigate the positional accuracy of a prototype X-ray imaging tool in combination with a real-time infrared tracking device allowing automated patient set-up in three dimensions.
MATERIAL AND METHODS: A prototype X-ray imaging tool has been integrated with a commercially released real-time infrared tracking device. The system, consisting of two X-ray tubes mounted to the ceiling and a centrally located amorphous silicon detector has been developed for automated patient positioning from outside the treatment room prior to treatment. Two major functions are supported: (a) automated fusion of the actual treatment images with digitally reconstructed radiographs (DRRs) representing the desired position; (b) matching of implanted radio opaque markers. Measurements of known translational (up to 30.0mm) and rotational (up to 4.0 degrees ) set-up errors in three dimensions as well as hidden target tests have been performed on anthropomorphic phantoms.
RESULTS: The system's accuracy can be represented with the mean three-dimensional displacement vector, which yielded 0.6mm (with an overall SD of 0.9mm) for the fusion of DRRs and X-ray images. Average deviations between known translational errors and calculations varied from -0.3 to 0.6mm with a standard deviation in the range of 0.6-1.2mm. The marker matching algorithm yielded a three-dimensional uncertainty of 0.3mm (overall SD: 0.4mm), with averages ranging from 0.0 to 0.3mm and a standard deviation in the range between 0.3 and 0.4mm.
CONCLUSIONS: The stereoscopic X-ray imaging device integrated with the real-time infrared tracking device represents a positioning tool allowing for the geometrical accuracy that is required for conformal radiation therapy of abdominal and pelvic lesions, within an acceptable time-frame.

Entities:  

Mesh:

Year:  2003        PMID: 12758249     DOI: 10.1016/s0167-8140(02)00385-7

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  29 in total

1.  Effect of image quality on correlation modeling error using a fiducial marker in a gimbaled linear accelerator.

Authors:  Hideharu Miura; Shuichi Ozawa; Tsubasa Enosaki; Fumika Hosono; Kiyoshi Yamada; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2019-02-25

Review 2.  Extracranial stereotactic body radiotherapy. Review of main SBRT features and indications in primary tumors.

Authors:  Carmen Rubio; Rosa Morera; Ovidio Hernando; Thomas Leroy; S Eric Lartigau
Journal:  Rep Pract Oncol Radiother       Date:  2013-11-01

3.  Quality and safety in stereotactic radiosurgery and stereotactic body radiation therapy: can more be done?

Authors:  Timothy D Solberg; Paul M Medin
Journal:  J Radiosurg SBRT       Date:  2011

4.  Radiosurgery technology development and use.

Authors:  Sanford L Meeks; Jason Pukala; Naren Ramakrishna; Twyla R Willoughby; Francis J Bova
Journal:  J Radiosurg SBRT       Date:  2011

5.  Image guided SBRT for multiple liver metastases with ExacTrac® Adaptive Gating.

Authors:  Carmen Rubio; Ovidio Hernando-Requejo; Daniel Zucca Aparicio; María ALlona Krauel; Mercedes López Gonzalez; Juan María Pérez; Emilio Sánchez Saugar; Pedro Fernández Letón
Journal:  Rep Pract Oncol Radiother       Date:  2016-09-17

6.  Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials.

Authors:  Reinhold Graf; Dirk Boehmer; Volker Budach; Peter Wust
Journal:  Strahlenther Onkol       Date:  2010-09-30       Impact factor: 3.621

7.  Spinal cord tolerance to single-fraction partial-volume irradiation: a swine model.

Authors:  Paul M Medin; Ryan D Foster; Albert J van der Kogel; James W Sayre; William H McBride; Timothy D Solberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-08       Impact factor: 7.038

8.  Reducing ExacTrac intrafraction imaging uncertainty for prostate stereotactic body radiotherapy using a pre-treatment CBCT.

Authors:  Barry Jordan; Luis Muñoz; Christopher Colyer
Journal:  Phys Eng Sci Med       Date:  2022-04-19

9.  Retroperitoneal tumour radiotherapy: clinical improvements using kilovoltage cone beam computed tomography.

Authors:  Xavier J Juan-Senabre; Carlos Ferrer-Albiach; Marta Rodríguez-Cordón; Agustín Santos-Serra; Juan López-Tarjuelo; Salvador Calzada-Feliu
Journal:  Clin Transl Oncol       Date:  2009-04       Impact factor: 3.405

10.  A closer look at the conventional Winston-Lutz test: Analysis in terms of dose.

Authors:  Juan-Francisco Calvo-Ortega; Sandra Moragues-Femenía; Coral Laosa-Bello; Sol San José-Maderuelo; Joan Casals-Farran
Journal:  Rep Pract Oncol Radiother       Date:  2019-07-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.