Total skin electron irradiation: evaluation of dose uniformity throughout the skin surface.

Yavuz Anacak1, Zumre Arican, Raquel Bar-Deroma, Ada Tamir, Abraham Kuten.   

Abstract

In this study, in vivo dosimetic data of 67 total skin electron irradiation (TSEI) treatments were analyzed. Thermoluminescent dosimetry (TLD) measurements were made at 10 different body points for every patient. The results demonstrated that the dose inhomogeneity throughout the skin surface is around 15%. The homogeneity was better at the trunk than at the extratrunk points, and was worse when a degrader was used. There was minimal improvement of homogeneity in subsequent days of treatment.

Entities:  

Mesh:

Year:  2003        PMID: 12747616     DOI: 10.1016/S0958-3947(02)00235-2

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  13 in total

Review 1.  Review of the results of the in vivo dosimetry during total skin electron beam therapy.

Authors:  Gabriele Guidi; Giovanni Gottardi; Paola Ceroni; Tiziana Costi
Journal:  Rep Pract Oncol Radiother       Date:  2013-08-15

2.  Time-gated scintillator imaging for real-time optical surface dosimetry in total skin electron therapy.

Authors:  Petr Bruza; Sarah L Gollub; Jacqueline M Andreozzi; Irwin I Tendler; Benjamin B Williams; Lesley A Jarvis; David J Gladstone; Brian W Pogue
Journal:  Phys Med Biol       Date:  2018-05-02       Impact factor: 3.609

3.  Helical irradiation of the total skin with dose painting to replace total skin electron beam therapy for therapy-refractory cutaneous CD4+ T-cell lymphoma.

Authors:  Chen-Hsi Hsieh; Pei-Wei Shueng; Shih-Chiang Lin; Hui-Ju Tien; An-Cheng Shiau; Yueh-Hung Chou; Meng-Hao Wu; Jen-Yu Wang; Chi-Kuan Chen; Yu-Jen Chen
Journal:  Biomed Res Int       Date:  2013-09-23       Impact factor: 3.411

4.  The dose penumbra of a custom-made shield used in hemibody skin electron irradiation.

Authors:  Charlotte I Rivers; Ismail AlDahlawi; Iris Z Wang; Anurag K Singh; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

5.  Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study.

Authors:  Alexander Nevelsky; Egor Borzov; Shahar Daniel; Raquel Bar-Deroma
Journal:  J Appl Clin Med Phys       Date:  2017-01       Impact factor: 2.102

6.  In vivo monitoring of total skin electron dose using optically stimulated luminescence dosimeters.

Authors:  Tanya Kairn; Rachael Wilks; Liting Yu; Craig Lancaster; Scott B Crowe
Journal:  Rep Pract Oncol Radiother       Date:  2019-12-16

7.  Dosimetric comparison of 4 MeV and 6 MeV electron beams for total skin irradiation.

Authors:  So-Yeon Park; Beom Seok Ahn; Jong Min Park; Sung-Joon Ye; Il Han Kim; Jung-In Kim
Journal:  Radiat Oncol       Date:  2014-09-06       Impact factor: 3.481

8.  Dosimetric characterization and optimization of a customized Stanford total skin electron irradiation (TSEI) technique.

Authors:  Felipe Luĉić; Beatriz Sánchez-Nieto; Paola Caprile; Gabriel Zelada; Karen Goset
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

9.  Total skin electron therapy in the lying-on-the-floor position using a customized flattening filter to eliminate field junctions.

Authors:  Christopher L Deufel; John A Antolak
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

10.  Validation of total skin electron irradiation (TSEI) technique dosimetry data by Monte Carlo simulation.

Authors:  Alexander Nevelsky; Egor Borzov; Shahar Daniel; Rachel Bar-Deroma
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.