Charlotte I Rivers1, Ismail AlDahlawi, Iris Z Wang, Anurag K Singh, Matthew B Podgorsak.
Abstract
We report our technique for hemibody skin electron irradiation with a custom-made plywood shield. The technique is similar to our clinical total skin electron irradiation (TSEI), performed with a six-pair dual field (Stanford technique) at an extended source-to-skin distance (SSD) of 377 cm, with the addition of a plywood shield placed at 50 cm from the patient. The shield is made of three layers of stan-dard 5/8'' thick plywood (total thickness of 4.75 cm) that are clamped securely on an adjustable-height stand. Gafchromic EBT3 films were used in assessing the shield's transmission factor and the extent of the dose penumbra region for two different shield-phantom gaps. The shield transmission factor was found to be about 10%. The width of the penumbra (80%-to-20% dose falloff) was measured to be 12 cm for a 50 cm shield-phantom gap, and reduced slightly to 10 cm for a 35 cm shield-phantom gap. In vivo dosimetry of a real case confirmed the expected shielded area dose.Entities:
Mesh:
Year: 2016 PMID: 27929500 PMCID: PMC5690530 DOI: 10.1120/jacmp.v17i6.6367
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Total skin electron irradiation (TSEI) beam dosimetric characteristics, for the combined six‐pair dual fields at an extended SSD of 377 cm
|
|
|
|---|---|
| Depth of maximum dose ( | 0.3 cm |
|
| 1.32 cm in water |
|
| 3.08 MeV |
|
| 2.21 cm in water |
| Bremsstrahlung photon dose background (at 3 cm depth) |
|
| Dose Rate Output at |
|
Figure 1A photograph showing the custom‐design plywood shield on its stand. Linac gantry is angled at 248°. Also shown behind the shield is the patient stand.
Figure 2Schematic of the film transmission and penumbra measurements setup.
Figure 3Gafchromic EBT3 film characteristic curve.
Figure 4Relative dose profile versus distance from the shield edge (positive distance is under shielded region) measured with Gafchromic EBT3 films for shield‐phantom gaps of 50 cm (squares) and 35 cm (triangles).
Results of in vivo dosimetry measurements for different body sites for two fractions of 180 cGy/fractions
|
|
|
|
|
| |
|---|---|---|---|---|---|
| Under Shield | Forehead | 0.03 – 0.03 | 20.5 – 20.5 | 20.5 | 0.11 |
| Upper Body | 0.02 – 0.03 | 13.5 – 20.5 | 17.6 | 0.10 | |
| Open | Lower Body (at level of pelvis) | 0.22 – 0.24 | 194.3 – 217.5 | 205.8 | 1.14 |
| Left anterior thigh | 0.23 – 0.24 | 205.8 – 217.5 | 212.8 | 1.18 | |
| Left ankle | 0.21 – 0.26 | 183.0 – 241.8 | 227.1 | 1.26 |