Electromagnetic interference by transcutaneous neuromuscular electrical stimulation in patients with bipolar sensing implantable cardioverter defibrillators: a pilot safety study.

Richard Crevenna1, Günter Stix, Johannes Pleiner, Thomas Pezawas, Herwig Schmidinger, Michael Quittan, Michael Wolzt.   

Abstract

Neuromuscular electrical stimulation (NMES) is a frequently applied therapy for the treatment of pain and a therapeutic option to increase thigh muscle strength and endurance capacity in patients with heart failure. Electromagnetic interference (EMI) by the signals with sensing of ICDs is possible. Eight patients with subpectoral ICD systems and different transvenous bipolar sensing leads were subjected to electrical stimulation of the neck and shoulder and of the thighs using different stimulation algorithms. EMI with ventricular sensing was detectable in three of eight subjects. EMI occurred during stimulation of the neck (n = 2) and thigh (n = 2). EMI by NMES with atrial sensing was seen in two of four subjects with dual chamber ICDs. The safety of peripheral NMES has to be individually tested as EMI can also occur in ICD patients with bipolar sensing.

Entities:  

Mesh:

Year:  2003        PMID: 12710324     DOI: 10.1046/j.1460-9592.2003.00104.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  6 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

2.  Effects of mobile telephones on the function of implantable cardioverter defibrillators.

Authors:  Izzet Tandogan; Bulent Ozin; Huseyin Bozbas; Sibel Turhan; Ramazan Ozdemir; Ertan Yetkin; Ergun Topal
Journal:  Ann Noninvasive Electrocardiol       Date:  2005-10       Impact factor: 1.468

3.  Inappropriate implantable cardioverter defibrillator shock from a transcutaneous muscle stimulation device therapy.

Authors:  Chung-Wah Siu; Hung-Fat Tse; Chu-Pak Lau
Journal:  J Interv Card Electrophysiol       Date:  2005-06       Impact factor: 1.900

4.  Lack of maintenance of shortwave diathermy equipment has a negative impact on power output.

Authors:  Rinaldo Roberto de Jesus Guirro; Elaine Caldeira de Oliveira Guirro; Natanael Teixeira Alves de Sousa
Journal:  J Phys Ther Sci       Date:  2014-04-23

Review 5.  Neuromuscular electrical stimulation of the thighs in cardiac patients with implantable cardioverter defibrillators.

Authors:  Fadime Cenik; Dieter Schoberwalter; Mohammad Keilani; Bruno Maehr; Michael Wolzt; Maximilian Marhold; Richard Crevenna
Journal:  Wien Klin Wochenschr       Date:  2016-07-25       Impact factor: 1.704

6.  The safety of electrical stimulation in patients with pacemakers and implantable cardioverter defibrillators: A systematic review.

Authors:  James Badger; Paul Taylor; Ian Swain
Journal:  J Rehabil Assist Technol Eng       Date:  2017-12-05