| Literature DB >> 27457876 |
Fadime Cenik1, Dieter Schoberwalter1,2, Mohammad Keilani1, Bruno Maehr1,3, Michael Wolzt4, Maximilian Marhold5, Richard Crevenna6.
Abstract
BACKGROUND: The aim of this systematic review was to update scientific knowledge concerning the safety of neuromuscular electrical stimulation (NMES) to increase exercise capacity and prevent cardiac cachexia in patients with implantable cardioverter defibrillators (ICDs).Entities:
Keywords: Electromagnetic interference; Heart failure; Implantable cardioverter defibrillator; Neuromuscular electrical stimulation; Risk
Mesh:
Year: 2016 PMID: 27457876 PMCID: PMC5104798 DOI: 10.1007/s00508-016-1045-2
Source DB: PubMed Journal: Wien Klin Wochenschr ISSN: 0043-5325 Impact factor: 1.704
Fig. 1Flowchart of the systematic literature research and the selection process
Summary of publications concerning neuromuscular electrical stimulation in cardiac patients with implantable cardioverter defibrillators
| Author | Study design | Implanted device | Type/location of stimulation | Stimulation parameters | Results |
|---|---|---|---|---|---|
| Kamiya et al. 2016 [ | Safety (pilot) study, ( | Left pectoral implanted ICD with dual-chamber lead system: | NMES of quadriceps and gastrocnemius muscles | 50 Hz, biphasic, 20 min, burst, 2.5 kHz, 5 s stim. +5 s interval, highest tolerable intensity (thigh: 25–60 mA; calf: 15–40 mA) | No EMI in all patients |
| Crevenna et al. 2004 [ | Safety study ( | Subpectoral implanted ICD: | NMES of thigh muscles | –4 patients: 63.3 Hz, 3.5 s on +4.5 s off, biphasic, 55–100 mA | No EMI in all patients |
| Crevenna et al. 2003 [ | Safety study ( | Subpectoral ICDs: | – IG 50, FM, HF TENS, LF TENS of trapezoid muscle | – IG50: dir. cur., 128 mA, 200 Hz, 400 μs stim. + serial duration of 50 ms | EMI caused by |
| Wayar et al. 2003 [ | Case report ( | 1st patient: pectoral ICD (Ventak® Mini III [Guidant Inc., St. Paul, Minnesota, USA]) | NMES of abdominal muscles | 55–75 Hz, biphasic, 7.3–10 mA, 5–11 V | EMI in both patients |
| 2nd patient: pectoral ICD (Ventak® AV [Guidant Inc., St. Paul, Minnesota, USA]) |
ICD implantable cardioverter defibrillator, NMES neuromuscular electrical stimulation, stim stimulation, EMI electromagnetic interference, IG50 Impulse galvanization, dir. cur. direct current, FM frequency modulation, HF-TENS high frequency transcutaneous electrical nerve stimulation, LF-TENS low frequency transcutaneous electrical nerve stimulation, FIB mimics tachycardia, thresh. dur. threshold duration, Stiwell 1200 (Med-El, Innsbruck, Austria) and Compex 2 (Compex SA, Ecublens, Switzerland) home devices