Nickel release after implantation of the Amplatzer occluder.

Marcus W Ries1, Christoph Kampmann, Hans-Jürgen Rupprecht, Gudrun Hintereder, Gerd Hafner, Jürgen Meyer.   

Abstract

BACKGROUND: Transcatheter closure of atrial septal defects is a new and less traumatic technique than open heart surgery. In recent years, patients with a patent foramen ovale sustaining potential paradoxical embolism have also become candidates for interventional closure devices. One of the more popular occluding devices is the Amplatzer septal occluder, which, like many other occluders, is made of nitinol. Nitinol-based alloys are widely used in medical products, for example, in orthopedics and orthodontics. However, the clinical use of nitinol, which contains 55% nickel, is still controversial because of concerns about its biocompatibility. Therefore, we examined the systemic nickel release after implantation of the Amplatzer occluder. METHODS AND
RESULTS: In 67 patients with no history of nickel sensitivity, blood samples were taken 24 hours before and 24 hours, 1, 3, and 12 months after occluder implantation. Nickel serum concentrations were measured by atomic absorption spectrometry; a value of <2 ng/mL of nickel was considered to be normal. A rise in mean serum levels of nickel was observed, from 0.47 ng/mL before implantation to 1.27 ng/mL (24 hours after), to a maximum of 1.50 ng/mL 1 month after implantation, which was statistically significant (P =.008 and P = 0.022, Wilcoxon Test). During follow-up, the values decreased to those measured before implantation.
CONCLUSIONS: Nickel seems to be released from the device, causing a systemic rise in serum levels of nickel, possibly until a calcium-phosphate layer has formed on the passive oxide film of the device or until endothelialization is complete. Possible biological effects should be considered, particularly in young patients or patients with nickel hypersensitivity.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12679773     DOI: 10.1067/mhj.2003.7

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  21 in total

1.  Reversible intracranial parenchymal changes in MRI after MCA aneurysm treatment with stent-assisted coiling technique; possible nickel allergy.

Authors:  Sıla Ulus; Abdullah Yakupoğlu; Ercan Kararslan; Civan Işlak; Aksel Siva; Naci Koçer
Journal:  Neuroradiology       Date:  2012-06-01       Impact factor: 2.804

Review 2.  Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.

Authors:  Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-06-23       Impact factor: 1.731

Review 3.  Systemic contact dermatitis and allergy to biomedical devices.

Authors:  Marcella Aquino; Tania Mucci
Journal:  Curr Allergy Asthma Rep       Date:  2013-10       Impact factor: 4.806

4.  Nickel-associated delayed multiple white matter lesions after stent-assisted coil embolization of intracranial unruptured aneurysm.

Authors:  Hun Soo Park; Ichiro Nakagawa; Shohei Yokoyama; Daisuke Wajima; Takeshi Wada; Yasushi Motoyama; Kimihiko Kichikawa; Hiroyuki Nakase
Journal:  BMJ Case Rep       Date:  2017-02-20

Review 5.  Nickel-free austenitic stainless steels for medical applications.

Authors:  Ke Yang; Yibin Ren
Journal:  Sci Technol Adv Mater       Date:  2010-02-26       Impact factor: 8.090

6.  Biocompatibility of new materials based on nano-structured nitinol with titanium and tantalum composite surface layers: experimental analysis in vitro and in vivo.

Authors:  Mikhail A Sevost'yanov; Elena O Nasakina; Alexander S Baikin; Konstantin V Sergienko; Sergey V Konushkin; Mikhail A Kaplan; Alexey V Seregin; Alexander V Leonov; Valery A Kozlov; Alexey V Shkirin; Nikolai F Bunkin; Alexey G Kolmakov; Sergey V Simakov; Sergey V Gudkov
Journal:  J Mater Sci Mater Med       Date:  2018-03-15       Impact factor: 3.896

7.  Endocarditis and Incomplete Endothelialization 12 Years after Amplatzer Septal Occluder Deployment.

Authors:  Allan K Nguyen; Brian A Palafox; Joanne P Starr; Richard N Gates; Farbouch Berdjis
Journal:  Tex Heart Inst J       Date:  2016-06-01

Review 8.  Biocompatibility screening in cardiovascular implants.

Authors:  M Sigler; T Paul; R G Grabitz
Journal:  Z Kardiol       Date:  2005-06

9.  Current practices in corrosion, surface characterization, and nickel leach testing of cardiovascular metallic implants.

Authors:  Srinidhi Nagaraja; Matthew Di Prima; David Saylor; Erica Takai
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-02-16       Impact factor: 3.368

Review 10.  Role of animal models for percutaneous atrial septal defect closure.

Authors:  Zakaria Jalal; Pierre-Emmanuel Seguela; Alban-Elouen Baruteau; David Benoist; Olivier Bernus; Olivier Villemain; Younes Boudjemline; Xavier Iriart; Jean-Benoit Thambo
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.