Validation of the Coulter LH 750 in a hospital reference laboratory.

C Aulesa1, I Pastor, D Naranjo, J Piqueras, R Galimany.   

Abstract

Validation of the Coulter LH 750 was carried out in our central hospital laboratory, which processes 1500 hemograms per day for patients admitted into the 4 hospitals of our hospital complex and their corresponding outpatient departments. It is the reference laboratory for the provincial health care area. The analytical quality and the practical use of the instrument were studied, and we obtained within- and between-run imprecision estimates with reference controls of between 0.24% and 2.56% and inaccuracies of between -1.32% and 3.07% for basic hemogram parameters. Similar results were obtained with specimens from patients, with imprecision estimates between 0.56% and 2.56%. Linearity estimates were between 0.3 x 10(9)/L and 380 x 10(9)/L for leukocytes and between 3 x 10(9)/L and 1900 x 10(9)/L for platelets. The instrument evaluation was completed with a study of interference by bilirubin, lipemia, hemolysis, platelet clumps, and heparin and an examination of other variables, including carryover, detection limits, and the correlation of results with those of the Coulter Gen-S. A special evaluation was made of the new erythroblast count feature; with reference controls, imprecision estimates for this count were 10% to 12%, and with patient specimens imprecision averaged 10.39% up to 4 erythroblasts per 100 leukocytes. We also studied the correlation of LH 750 results and interferences with those of the manual reference method. Finally, National Committee for Clinical Laboratory Standards protocols were used to test for suspect and confirmation flags in leukocyte differential counts for 258 specimens representative of our routine. The practicability of the analyzer was studied in terms of technical difficulty, speed, and cost; also evaluated were new software elements for validation by source clinic and pathology and for reference values based on age. In conclusion, we analyzed the impact and improvements that may be expected in our laboratory (a user of Technicon and Coulter instruments) from introducing the new LH 750 analyzer into our routine.

Entities:  

Mesh:

Year:  2003        PMID: 12661824

Source DB:  PubMed          Journal:  Lab Hematol        ISSN: 1080-2924


  5 in total

1.  Three-way comparison of methods for the measurement of the erythrocyte sedimentation rate.

Authors:  Johnny Ndoni Mahlangu; Melony Davids
Journal:  J Clin Lab Anal       Date:  2008       Impact factor: 2.352

2.  Properties of human blood monocytes. I. CD91 expression and log orthogonal light scatter provide a robust method to identify monocytes that is more accurate than CD14 expression.

Authors:  Dorothy Hudig; Kenneth W Hunter; W John Diamond; Doug Redelman
Journal:  Cytometry B Clin Cytom       Date:  2013-10-08       Impact factor: 3.058

3.  CellCODE: a robust latent variable approach to differential expression analysis for heterogeneous cell populations.

Authors:  Maria Chikina; Elena Zaslavsky; Stuart C Sealfon
Journal:  Bioinformatics       Date:  2015-01-11       Impact factor: 6.937

4.  Initial performance evaluation of the UniCel® DxH 800 Coulter® cellular analysis system.

Authors:  B D Hedley; M Keeney; I Chin-Yee; W Brown
Journal:  Int J Lab Hematol       Date:  2011-02       Impact factor: 2.877

5.  Radiofrequency radiation and human ferritin.

Authors:  Viroj Wiwanitkit
Journal:  J Med Signals Sens       Date:  2013-01
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.