Performance evaluation of a multi-slice CT system with 16-slice detector and increased gantry rotation speed for isotropic submillimeter imaging of the heart.

Thomas Flohr1, Axel Küttner, Herbert Bruder, Karl Stierstorfer, Sandra S Halliburton, Stefan Schaller, Bernd M Ohnesorge.   

Abstract

BACKGROUND: 4-slice CT scanners have shown limitations in clinical application for noninvasive coronary CT angiography (CTA). We evaluate advances in ECG-gated scanning of the heart and the coronary arteries with recently introduced 16-slice CT equipment (SOMATOM Sensation 16, Siemens, Forchheim, Germany).
MATERIALS AND METHODS: The technical principles of ECG-gated cardiac scanning, scan parameters, and detector design of the new scanner are presented. ECG-gated scan and image reconstruction techniques and ECG-controlled dose modulation ("ECG pulsing") for a reduction of the patient dose are described, key parameters for image quality and simulation results presented, and phantom studies and initial patient experience discussed. The impact of reduced gantry rotation time (0.42 s) on temporal resolution and initial estimations of the patient dose are presented.
RESULTS: Extensions of ECG-gated reconstruction algorithms used for 4-slice CT provide adequate image quality for up to 16 slices. For each detector collimation different slice widths are available for retrospective reconstruction with well-defined slice sensitivity profiles (SSPs). For coronary CTA the heart can be covered with 0.75 mm collimation within a 20-s breathhold. The best possible spatial resolution is 0.5 x 0.5 x 0.6 mm. For 0.42 s gantry rotation time, temporal resolution reaches its optimum (105 ms) at a heart rate of 81 bpm. Effective patient dose for coronary CTA is 4-5 mSv using ECG-pulsed acquisition.
CONCLUSION: The clinical performance of coronary CTA by means of spatial resolution, temporal resolution and scan time is substantially improved with the evaluated 16-slice CT scanner. Also, display of smaller coronary segments and instent visualization are substantially improved.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 12616316     DOI: 10.1007/s00059-003-2456-1

Source DB:  PubMed          Journal:  Herz        ISSN: 0340-9937            Impact factor:   1.443


  6 in total

Review 1.  Multidetector-row CT: cardiosurgery indications.

Authors:  Christopher Herzog; Selami Dogan; Gerhard Wimmer-Greinecker; Joern O Balzer; Martin G Mack; Thomas J Vogl
Journal:  Eur Radiol       Date:  2003-12       Impact factor: 5.315

2.  Evaluation of spatial and temporal resolution for ECG-gated 16-row multidetector CT using a dynamic cardiac phantom.

Authors:  P G C Begemann; U van Stevendaal; R Manzke; A Stork; F Weiss; C Nolte-Ernsting; M Grass; G Adam
Journal:  Eur Radiol       Date:  2005-01-21       Impact factor: 5.315

3.  Non-invasive coronary angiography: the clinical value of multi-slice computed tomography in the assessment of patients with prior coronary bypass surgery. Evaluating grafts and native vessels.

Authors:  Helge von Kiedrowski; Marcus Wiemer; Krista Franzke; Rainer Preuss; Bernhard Vaske; Thomas Butz; Olaf Oldenburg; Thomas Bitter; Khalid Mahmood; Wolfram Burchert; Dieter Horstkotte; Christoph Langer
Journal:  Int J Cardiovasc Imaging       Date:  2008-09-01       Impact factor: 2.357

4.  Italian multicenter, prospective study to evaluate the negative predictive value of 16- and 64-slice MDCT imaging in patients scheduled for coronary angiography (NIMISCAD-Non Invasive Multicenter Italian Study for Coronary Artery Disease).

Authors:  Riccardo Marano; Francesco De Cobelli; Irene Floriani; Christoph Becker; Christopher Herzog; Maurizio Centonze; Giovanni Morana; Gian Franco Gualdi; Guido Ligabue; Gianluca Pontone; Carlo Catalano; Dante Chiappino; Massimo Midiri; Giovanni Simonetti; Filippo Marchisio; Lucio Olivetti; Rossella Fattori; Lorenzo Bonomo; Alessandro Del Maschio
Journal:  Eur Radiol       Date:  2008-12-17       Impact factor: 5.315

5.  A new approach to the assessment of lumen visibility of coronary artery stent at various heart rates using 64-slice MDCT.

Authors:  J M Groen; M J W Greuter; P M A van Ooijen; M Oudkerk
Journal:  Eur Radiol       Date:  2007-02-16       Impact factor: 5.315

6.  Assessment of hemodynamics in a rat model of liver cirrhosis with precancerous lesions using multislice spiral CT perfusion imaging.

Authors:  Guolin Ma; Rongjie Bai; Huijie Jiang; Xuejia Hao; Zaisheng Ling; Kefeng Li
Journal:  Biomed Res Int       Date:  2013-06-20       Impact factor: 3.411

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.