Non-invasive estimation of cardiac output in critical care patients.

U M Gerhardt1, C Schöller, D Böcker, H Hohage.   

Abstract

OBJECTIVE: This study was carried out to compare cardiac output measurements determined by thermodilution and by Portapres, a non-invasive system. DESIGN, PATIENTS AND
SETTING: Eighty-seven non-invasive blood pressure measurements were performed in 46 patients in our critical care unit utilising the new, non-invasive Portapres system. Cardiac output values were obtained from these blood pressure values using an aortic impedance model and compared to cardiac output values estimated by the thermodilution technique.
MEASUREMENTS AND MAIN RESULTS: Statistically significant (p < 0.01) differences (2.3 l/min; limits of agreement +/-5 l/min) were noted between invasive and non-invasive cardiac output measurements. Differences in measured cardiac outputs increased for patients receiving catecholamine therapy, in patients with hemodynamic instability (e.g., sepsis and cardiac insufficiency), in patients with artificial ventilation, in patients with long duration of intensive care, in younger (<60 yr) patients and in women. We found no influence of the body mass index (BMI) on the accuracy of Portapres results. In only one single subgroup, 10 patients with pulmonary diseases, Portapres measurements were not statistically significant different from reference results.
CONCLUSIONS: To date, Portapres measurements cannot replace thermodilution cardiac output estimations. Fluctuations of finger arterial perfusion due to hemodynamic instability, hypothermia and catecholamines may be responsible for problems of Portapres use in critically ill patients.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 12578073     DOI: 10.1023/a:1011481115195

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  25 in total

1.  Continuous cardiac output in septic shock by simulating a model of the aortic input impedance: a comparison with bolus injection thermodilution.

Authors:  W T Jellema; K H Wesseling; A B Groeneveld; C P Stoutenbeek; L G Thijs; J J van Lieshout
Journal:  Anesthesiology       Date:  1999-05       Impact factor: 7.892

2.  Non-invasive continuous finger blood pressure measurement during orthostatic stress compared to intra-arterial pressure.

Authors:  B P Imholz; J J Settels; A H van der Meiracker; K H Wesseling; W Wieling
Journal:  Cardiovasc Res       Date:  1990-03       Impact factor: 10.787

3.  Measurement of cardiac output by transesophageal echocardiography in mechanically ventilated patients. Comparison with thermodilution.

Authors:  P Estagnasié; K Djedaini; L Mier; F Coste; D Dreyfuss
Journal:  Intensive Care Med       Date:  1997-07       Impact factor: 17.440

4.  Disparities between aortic and peripheral pulse pressures induced by upright exercise and vasomotor changes in man.

Authors:  L B Rowell; G L Brengelmann; J R Blackmon; R A Bruce; J A Murray
Journal:  Circulation       Date:  1968-06       Impact factor: 29.690

5.  Continuous non-invasive blood pressure monitoring: reliability of Finapres device during the Valsalva manoeuvre.

Authors:  B P Imholz; G A van Montfrans; J J Settels; G M van der Hoeven; J M Karemaker; W Wieling
Journal:  Cardiovasc Res       Date:  1988-06       Impact factor: 10.787

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  A comparison of non-invasive continuous finger blood pressure measurement (Finapres) with intra-arterial pressure during prolonged head-up tilt.

Authors:  M E Petersen; T R Williams; R Sutton
Journal:  Eur Heart J       Date:  1995-11       Impact factor: 29.983

8.  Temporal changes in left ventricular function after massive sympathetic nervous system activation.

Authors:  S A Lang; M B Maron; F J Bosso; C F Pilati
Journal:  Can J Physiol Pharmacol       Date:  1994-06       Impact factor: 2.273

9.  Measurement of cardiac output by pulse dye densitometry using indocyanine green: a comparison with the thermodilution method.

Authors:  T Imai; K Takahashi; H Fukura; Y Morishita
Journal:  Anesthesiology       Date:  1997-10       Impact factor: 7.892

10.  Noninvasive assessment of cardiac output in critically ill patients by analysis of the finger blood pressure waveform.

Authors:  M M Hirschl; M Binder; M Gwechenberger; H Herkner; A Bur; H Kittler; A N Laggner
Journal:  Crit Care Med       Date:  1997-11       Impact factor: 7.598

View more
  1 in total

1.  Noninvasive cardiac output monitoring during exercise testing: Nexfin pulse contour analysis compared to an inert gas rebreathing method and respired gas analysis.

Authors:  Sebastiaan A Bartels; Wim J Stok; Rick Bezemer; Remco J Boksem; Jeroen van Goudoever; Thomas G V Cherpanath; Johannes J van Lieshout; Berend E Westerhof; John M Karemaker; Can Ince
Journal:  J Clin Monit Comput       Date:  2011-10-15       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.