Ergonomic evaluation of an ecological interface and a profilogram display for hemodynamic monitoring.

A Jungk1, B Thull, A Hoeft, G Rau.   

Abstract

OBJECTIVE: Comprehensive monitoring of the patient state and subsequent decision making is an essential part of the task of an anaesthetist. The physicians' decision making process is based upon a concept of partly abstract physiologic parameters such as depth of anaesthesia or contractility. This concept is derived from the measured parameters given on todays' trend displays in addition to context information available for the anaesthetist. We investigated two alternative approaches of display design for hemodynamic monitoring: 1) integrated displays based on ecological interface design, and 2) profilogram displays based on intelligent alarms.
METHOD: To evaluate differences in decision making, the two displays and a trend display were compared in an experimental set-up with computer simulated vital parameter curves. From a start state with random parameter deviations from the ideal state, subjects had to achieve the ideal circulatory performance as fast as possible by manipulating vasomotor tone, heart rate, blood volume and contractility. To analyse subjects' decision making process, eye-tracking, event-logging, and the method of think aloud protocols were used. Twenty anaesthesiologists performed 113 experiments (approximately 2 with each display).
RESULTS: The anaesthetists failed to achieve the task in 37% using the trend display, in 19% using the profilogram display, and in 13% using the ecological interface. Hence, a safer task solution was possible with the ecological interface and the profilogram display but at the expense of various performance parameters such as higher trial time, more interactions with the simulated system, and more frequent eye movements. In contrast to the trend display and the profilogram display, where anaesthetists were mainly focussed on controlling the left atrial pressure, such an behaviour was less observed with the ecological interface.
CONCLUSION: Our results have shown that subjects came to more effective solutions with the traditional trend display. The main reason for this result may be their years of experience with this kind of display type. Regarding safe and goal-intended decision finding, the results are encouraging for further experiments with redesigned ecological displays. But these displays ought to have smoother changes with respect to the traditional trend displays. Furthermore, new experiments have to be performed under real or fairly real (e.g. together with an anaesthesia simulator) conditions to underline the positive results for ecological interfaces.

Entities:  

Mesh:

Year:  1999        PMID: 12578045     DOI: 10.1023/a:1009909229827

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  19 in total

Review 1.  Graphical displays: implications for divided attention, focused attention, and problem solving.

Authors:  K B Bennett; J M Flach
Journal:  Hum Factors       Date:  1992-10       Impact factor: 2.888

2.  Incorporating user and dialogue models into the interface design of an intelligent patient monitor.

Authors:  E Coiera
Journal:  Med Inform (Lond)       Date:  1991 Oct-Dec

3.  Accidents, near accidents and complications during anaesthesia. A retrospective analysis of a 10-year period in a teaching hospital.

Authors:  V Chopra; J G Bovill; J Spierdijk
Journal:  Anaesthesia       Date:  1990-01       Impact factor: 6.955

Review 4.  [Status of eye movement data in the microanalysis of cognitive processes].

Authors:  H W Schroiff
Journal:  Z Psychol Z Angew Psychol       Date:  1987

5.  Decision making in high dependency environments--can we learn from modern industrial management models?

Authors:  W Friesdorf; S Konichezky; F Gross-Alltag; D Geva; M Nathe; S Schraag
Journal:  Int J Clin Monit Comput       Date:  1994-02

Review 6.  Human error in anesthetic mishaps.

Authors:  D M Gaba
Journal:  Int Anesthesiol Clin       Date:  1989

7.  A circulatory model for combined nitroprusside-dopamine therapy in acute heart failure.

Authors:  C Yu; R J Roy; H Kaufman
Journal:  Med Prog Technol       Date:  1990-05

8.  The Australian Incident Monitoring Study: an analysis of 2000 incident reports.

Authors:  R K Webb; M Currie; C A Morgan; J A Williamson; P Mackay; W J Russell; W B Runciman
Journal:  Anaesth Intensive Care       Date:  1993-10       Impact factor: 1.669

9.  An analysis of major errors and equipment failures in anesthesia management: considerations for prevention and detection.

Authors:  J B Cooper; R S Newbower; R J Kitz
Journal:  Anesthesiology       Date:  1984-01       Impact factor: 7.892

10.  Anesthetic mishaps: breaking the chain of accident evolution.

Authors:  D M Gaba; M Maxwell; A DeAnda
Journal:  Anesthesiology       Date:  1987-05       Impact factor: 7.892

View more
  7 in total

1.  The employment of an iterative design process to develop a pulmonary graphical display.

Authors:  S Blake Wachter; Jim Agutter; Noah Syroid; Frank Drews; Matthew B Weinger; Dwayne Westenskow
Journal:  J Am Med Inform Assoc       Date:  2003-03-28       Impact factor: 4.497

2.  Evaluation of two new ecological interface approaches for the anesthesia workplace.

Authors:  A Jungk; B Thull; A Hoeft; G Rau
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

Review 3.  Evaluations of physiological monitoring displays: a systematic review.

Authors:  Matthias Görges; Nancy Staggers
Journal:  J Clin Monit Comput       Date:  2007-12-07       Impact factor: 2.502

Review 4.  Novel Interface Designs for Patient Monitoring Applications in Critical Care Medicine: Human Factors Review.

Authors:  Evismar Andrade; Leo Quinlan; Richard Harte; Dara Byrne; Enda Fallon; Martina Kelly; Siobhan Casey; Frank Kirrane; Paul O'Connor; Denis O'Hora; Michael Scully; John Laffey; Patrick Pladys; Alain Beuchée; Gearóid ÓLaighin
Journal:  JMIR Hum Factors       Date:  2020-07-03

5.  Critical care information display approaches and design frameworks: A systematic review and meta-analysis.

Authors:  Melanie C Wright; Damian Borbolla; Rosalie G Waller; Guilherme Del Fiol; Thomas Reese; Paige Nesbitt; Noa Segall
Journal:  J Biomed Inform X       Date:  2019-06-22

6.  Avatar-based versus conventional vital sign display in a central monitor for monitoring multiple patients: a multicenter computer-based laboratory study.

Authors:  Olivier Garot; Julian Rössler; Juliane Pfarr; Michael T Ganter; Donat R Spahn; Christoph B Nöthiger; David W Tscholl
Journal:  BMC Med Inform Decis Mak       Date:  2020-02-10       Impact factor: 2.796

Review 7.  Situation Awareness-Oriented Patient Monitoring with Visual Patient Technology: A Qualitative Review of the Primary Research.

Authors:  David Werner Tscholl; Julian Rössler; Sadiq Said; Alexander Kaserer; Donat Rudolf Spahn; Christoph Beat Nöthiger
Journal:  Sensors (Basel)       Date:  2020-04-09       Impact factor: 3.576

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.